8 Service routines
8.1 Perform service routines via the control panel
310
In the test mode, travel to the center position of the ceph scan
sensor axis is initially performed without offsets. The T key lights
up and indicates the test mode.
5. Move the sensor to its front position by touching the R key (B). The
sensor can always be moved back and forth between its front and
center positions with the R key.
6. Move the sensor to its center position by touching the R key.
7. Measure distances L2 and L1 with a slide gage.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016