Sirona Dental Systems GmbH
Service Manual GALILEOS
61 25 665 D3437
D3437.076.01.15.02 06.2012
7.3
Adjustment and calibration via the "Service
Functions" menu
Adjustment and calibration via the "Service Functions" menu
7.3.1
Diaphragm image
Diaphragm image
7.3.1.1
"Type 1/Type 2" diaphragm
"Type 1/Type 2" diaphragm
✔
"Service functions"
The
1. Click on the
SIDEXIS XG makes the unit ready for exposure [ → 162].
Service routine S030.5 is displayed on the control panel.
2. Take an exposure (85 kV/21 mAs) [ → 162].
For "GALILEOS Comfort":
3.
7.3 Adjustment and calibration via the "Service Functions" menu
menu is selected.
"Image acquisition"
button.
Evaluate the image.
7 Adjusting/calibrating the unit