Symbol description ..................Abbreviations ....................Safety information ....................Technical description....................Task ......................Functionality ....................SIROInspect probe design ................Scope of supply..................... Technical data....................Service life of Sirona instruments..............Operation........................Connect the power supply unit..............Replace and charge the battery ..............
Page 3
Sirona Dental Systems GmbH Operating Instructions SIROInspect Working clearances..................Storage and transport conditions................Disposal........................63 82 936 D3584 D3584.201.01.03.02 02.2013...
SIROInspect aids detection and monitoring of carious tooth substance during excavation of previously opened cavities. Contraindications Do not use SIROInspect in conjunction with caries detection dyes! Do not use SIROInspect if you have red-green color blindness or visual impairment. Target group Trained dental personnel This product is intended for use only by trained dental personnel in dental practices and laboratories.
Sirona Dental Systems GmbH 1 Before you begin … Operating Instructions SIROInspect 1.2 Formats and symbols used Formats and symbols used Formats and symbols used The formats and symbols used in this document have the following meaning: ✔ Requests you to do something.
Blinding Do not shine directly into the eyes. Only switch the SIROInspect on once it is in the patient’s mouth to avoid glare. The glare is not harmful, but may temporarily impair vision. UV radiation LED product of risk class 0. SIROInspect emits UV light. Prevent the oral mucosa from being unnecessarily exposed to radiation from UV rays.
3 Technical description Sirona Dental Systems GmbH 3.1 Task Operating Instructions SIROInspect Technical description Technical description Task Task SIROInspect ® FACE (Fluorescence aided caries excavation) is a method which enables the practitioner to remove infected tooth material in a targeted manner, based on the special fluorescent properties of the decomposition products resulting from caries-causing bacteria.
SIROInspect 90° light guide E SIROInspect attachable filter SIROInspect base body SIROInspect power supply unit Scope of supply Scope of supply Basic equipment, SIROInspect diagnostic set, REF 63 25 174 Quantity Designation REF No. SIROInspect base body 63 75 401 SIROInspect 90°...
Sirona Dental Systems GmbH 3 Technical description Operating Instructions SIROInspect 3.6 Service life of Sirona instruments Service life of Sirona instruments Service life of Sirona instruments Service life 5 + 3 years /3 + 1 1/2 years When used as intended: Additional service life periods_FACE ●...
➢ Hold the battery as far back as possible. 1. Remove the discharged battery (B) from SIROInspect (A). 2. Insert the charged battery into SIROInspect until it audibly clicks into place. 3. Position the discharged battery onto the charging station (C) until it audibly clicks into place.
Sirona Dental Systems GmbH 4 Operation Operating Instructions SIROInspect 4.3 Install attachable filter Install attachable filter ➢ Slide the attachable filter from the front onto the base body. Press the Install attachable filter silicone ring with the fingers until the attachable filter is firmly in place.
SIROInspect cannot assess initial caries. Use SIROInspect if an initial diagnosis has been made and a decision has already been made on how the caries is to be treated and if the cavity is open.
Manual cleaning must always be combined with disinfection. Cleaning SIROInspect The SIROInspect must not be cleaned or disinfected with a machine. 1. Remove the silicone ring from the attachable filter. 2. Clean the light guide, the silicone ring and attachable filter by placing it under running water (<...
Use only disinfectants that comply with the requirements of your national authorities and whose bactericidal, fungicidal and virucidal properties have been tested and properly certified. Sirona recommends using MinuteWipes from Alpro. In the US: ™ Caviwipes Observe the instructions of the disinfectant manufacturers.
6 Electromagnetic compatibility Operating Instructions SIROInspect 6.1 Electromagnetic emission Electromagnetic compatibility Electromagnetic compatibility NOTICE SIROInspect complies with all requirements for electromagnetic compatibility according to IEC 60601-1-2: 2007. Definitions: Emission (electromagnetic) When electromagnetic energy is emitted by a source. Interference immunity The ability of a device or system to work without errors even if there is electromagnetic interference.
6 Electromagnetic compatibility Sirona Dental Systems GmbH 6.2 Interference immunity Operating Instructions SIROInspect Interference immunity Interference immunity The UNIT is intended for operation in the electromagnetic environment specified below. The customer or user of the UNIT should make sure that it is used in such an environment.
Page 20
6 Electromagnetic compatibility Sirona Dental Systems GmbH 6.3 Working clearances Operating Instructions SIROInspect Rated maximum output power of Working clearance according to transmission frequency [m] transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz d= [1.2] √P...
Sirona Dental Systems GmbH 7 Storage and transport conditions Operating Instructions SIROInspect Storage and transport conditions Storage and transport conditions Dispatch Clean and disinfect the instrument or accessory used on the patient before sending it in. Protect against Sensitive moisture...
Disposing of batteries and charging stations Batteries must be disposed of according to disposal regulations and statutory provisions in your country. Remove the battery from the base body of the SIROInspect probe before disposal. Your product is marked with the adjacent symbol. Within the European Economic Area, this product is subject to Directive 2002/96/EC as well as the corresponding national laws.
Page 23
Sirona Dental Systems GmbH 8 Disposal Operating Instructions SIROInspect All disassembly, transport and packaging costs are to be borne by the owner / operator of the equipment. The disposal itself is free of charge. Worldwide (outside Germany): Please contact your local dental equipment specialist for country-specific information on disposal.