Troubleshooting; Error Logging Memory - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
6

Troubleshooting

DANGER
Potentially lethal shock hazard!
It is essential to switch off the unit and wait at least 1 minute before
removing a cover.
Switch OFF the X-ray unit before connecting a measuring instrument.
Perform continuity tests only on units which are switched off.
NOTICE
Risk of damage to unit
Select the correct current/voltage type and adjust the measuring range
to match the expected readings.
Keep to the prescribed cool-off periods if several exposures have to be
taken to check a measured value.
Please observe the usual precautionary measures for handling printed
circuit boards (ESD). Touch a ground point to discharge static electricity
before touching any boards.
CAN bus cable: When unplugging CAN bus cables, it is essential to
unplug the power supply as well.
6.1

Error logging memory

The error logging memory is part of the
entitled "Opening / saving "Unit information" [ → 59]").
Data which might be expected to occur in the error logging memory is
explained below to aid you in interpreting it.
6 Troubleshooting
6.1 Error logging memory
"Device details"
(see the section
139