S002: Radiation Without Rotary Movement, Selectable Kv/Ma Level 255 And Maximum Radiation Time 8.1.5.1 S002: Test Step 1 - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
S002
S002
S002.1
S002.3
S002.4
64 95 258 D3632
D3632.076.01.03.02 09.2016
8.1.5
S002: Radiation without rotary movement, selectable kV/
mA level and maximum radiation time
SA**
Function
X-ray beam test
Explanation:
● Radiation with selected kV/mA level
● Maximum radiation time selectable
● Radiation from any position
yes
The primary diaphragm remains in the last position selected.
yes
The primary diaphragm is opened fully.
yes
The primary diaphragm is set to a widened opening symmetrical to the
PAN setting so that the step filter still remains covered.
* SR=service routine, ** SA=security access
8.1.5.1
S002: Test step 1
Radiation without rotary movement; selectable kV/mA level and
maximum radiation time; primary diaphragm in last position selected
Unit is radiating X-rays
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
Selection field
1
2
Factory settings: 60kV/8mA, 2s
1. Call service routine S002.1 [ → 249].
2. Use the arrow keys (A) to select the required kV/mA level and the
required radiation time (see table).
8.1 Perform service routines via the control panel
WARNING
Parameters
kV/mA level
Radiation time
8 Service routines
Range of values
60 kV/3mA –
90kV/12mA –
0.1s – 4s
255