Service Manual ORTHOPHOS SL
Error code
Description
E6 14 06
The bootline of board DX11 had to
be preassigned with default values.
Error code
Description
E6 14 10
Clock signals for sensor image
transfer not received on board DX1/
DX11 (...10).
Faulty detection of sensor image
E6 14 11
transfer data signals on board DX1/
DX11; nonrecurring (...11).
Faulty detection of sensor image
transfer data signals on board DX1/
E6 14 12
DX11; recurring (...12).
64 95 258 D3632
D3632.076.01.03.02 09.2016
Actions required
E6 14 06
● Re-configuration of network data required.
Actions required
E6 14 10/11/12 Pan/Ceph
● Acknowledge error and, if possible, check
function at another slot (pan/ceph) via test
image (ceph).
If both test images indicate the error ...
● Check image signal on board DX1 (does
the image LED on DX1 switch on when the
ring cycle starts in test image mode?),
replace DX1 if necessary.
If only the ceph test image is not OK...
For E6 14 10 and ceph exposure:
● Check board DX91 and cables L34, L40,
replace if necessary.
● Check 2D data path for ceph via service
routine S033.10 (2D test image) and
replace sensor or cable if necessary.
Otherwise:
● Check cables L35, L38 for crushed spots
and kinks and check connectors, repair or
replace if necessary.
● Check board DX1, replace if necessary.
● Run service routine S033.10. This
determines whether the fault is stationary
or permanent. Replace sensor or cable if
necessary.
5 Messages
5.4 List of error messages
see
see
S. [ → 396]
S. [ → 146],
S. [ → 153],
S. [ → 307],
S. [ → 404],
S. [ → 396]