9 Repair
9.12 Sensor unit
378
9.12

Sensor unit

NOTICE
Damage to the Flat Panel Detector
The Flat Panel Detector is a sensitive component. Be careful not to
touch the detector surface during assembly.
NOTICE
Damage to the rotation unit
The sensor is rotated via a motor drive. The gearing of the rotation unit
can be damaged if it is turned by hand.
9.12.1

Flat Panel Detector

9.12.1.1

Replacing the flat panel detector

1. Loosen the screws (F).
2. Release the 4 latches on the sensor cover (D) and carefully remove
the covers (D and E) from the sensor unit.
3. Pull the cable L84 off from board DX83.