Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Error code
Description
E6 61 06
Module failed in TTP (detected on
master side)
TTP = Time Trigger Protocol
Error code
Description
E6 61 07
TTP timeout error (detected on
slave side).
The module was temporarily not
addressed by the master:
Undervoltage processing error in
the software on the master side.
Master (DX11) receives no return
commanding from the module
This error may also occur in
connection with other causal error
messages! Please also observe the
causal error message! It appears
only after you acknowledge the first
error message.
TTP = Time Trigger Protocol
Error code
Description
E6 61 08
General fault detected locally on
module (slave side). CAN controller
being reinitialized.
64 95 258 D3632
D3632.076.01.03.02 09.2016
Actions required
E6 61 06
● Check the CAN bus.
● Check board DX61; replace the
diaphragm unit, if necessary.
● Please contact the Sirona Customer
Service Center (CSC) to find out whether
a bug fix by means of a software update is
possible and perform such an update if
necessary.
Actions required
E6 61 07
● Check the CAN bus.
● Check power supply of board DX11;
measuring point 3.3 V on board DX1 (see
wiring diagrams).
If 3.3 V present...
● Replace board DX11.
If 3.3 V not present...
● Replace board DX1.
Actions required
E6 61 08
● Check the CAN bus.
● Check software versions on the info
screen or by running service routine
S008.2, run software update if necessary.
● Check board DX61; replace the
diaphragm unit, if necessary.
● Please contact the Sirona Customer
Service Center (CSC) to find out whether
a bug fix by means of a software update is
possible and perform such an update if
necessary.
5 Messages
5.4 List of error messages
see
S. [ → 141],
S. [ → 146],
S. [ → 370],
S. [ → 68]
see
S. [ → 141]
S. [ → 397]
S. [ → 396]
see
S. [ → 141],
S. [ → 267],
S. [ → 146],
S. [ → 370],
S. [ → 370]
125