Sirona ORTHOPHOS SL 2D Service Manual page 211

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Lx
Ly
D
64 95 258 D3632
D3632.076.01.03.02 09.2016
3. Set the diaphragm tilt via screw (C) (Lx in mm) and the diaphragm
height via screw (D) (Ly in mm).
Screw (C):
Turn counterclockwise = adjusts the diaphragm to the right
Turn clockwise = adjusts the diaphragm to the left
Screw (D):
Turn counterclockwise = adjusts the diaphragm downward
Turn clockwise = adjusts the diaphragm upward
To measure the adjustment, use a gauge.
The above-mentioned directions of rotation for the screws apply to
both left-arm and right-arm versions.
4. Retighten screws (B) firmly.
5. Once adjustment is complete, put the secondary diaphragm cover
C
back again and fix it in place with the screw (A).
7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
211