Sirona ORTHOPHOS SL 2D Service Manual
Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

kÉï=~ë=çÑW=
loqelmelp=pi=Oa=L=loqelmelp=pi=Oa=`ÉéÜ
loqelmelp=pi=Pa=L=loqelmelp=pi=Pa=`ÉéÜ
pÉêîáÅÉ=j~åì~ä
MVKOMNS
ORTHOPHOS SL / Ceph Service Manual
=
båÖäáëÜ
båÖäáëÜ

Advertisement

Table of Contents
loading

Summary of Contents for Sirona ORTHOPHOS SL 2D

  • Page 1 MVKOMNS kÉï=~ë=çÑW= loqelmelp=pi=Oa=L=loqelmelp=pi=Oa=`ÉéÜ loqelmelp=pi=Pa=L=loqelmelp=pi=Pa=`ÉéÜ pÉêîáÅÉ=j~åì~ä båÖäáëÜ båÖäáëÜ ORTHOPHOS SL / Ceph Service Manual...
  • Page 3: Table Of Contents

    Sirona Dental Systems GmbH Table of contents Service Manual ORTHOPHOS SL Table of contents About this Service Manual..................Other relevant documents ................Scope ......................Tools and auxiliary materials................. Structure of the document................1.4.1 Identification of the danger levels............. 1.4.2 Formats and symbols used ..............
  • Page 4 Table of contents Sirona Dental Systems GmbH Service Manual ORTHOPHOS SL 3.1.7 Board photos..................3.1.7.1 Boards in the slide ............3.1.7.2 Boards in the stand............3.1.7.3 Boards in the cephalometer (left- and right-arm versions) 3.1.7.4 Board in the remote control ..........
  • Page 5 Sirona Dental Systems GmbH Table of contents Service Manual ORTHOPHOS SL List of error messages.................. 5.4.1 Location 06: Tube assembly/DX6 ........... 5.4.2 Location 07: Easypad/DX7.............. 5.4.3 Location 10: System hardware............5.4.4 Location 11: Power PC/Board DX11 ..........5.4.5 Location 12: CAN bus ..............
  • Page 6 Table of contents Sirona Dental Systems GmbH Service Manual ORTHOPHOS SL 7.1.5 Save values..................7.1.6 Test phantom for adjustment and calibration ........7.1.6.1 Needle phantom for panoramic adjustment....7.1.6.2 Adjustment phantom for adjustment of the cephalometer 178 7.1.6.3 Geometry phantom for volume calibration.....
  • Page 7 Sirona Dental Systems GmbH Table of contents Service Manual ORTHOPHOS SL 8.1.5 S002: Radiation without rotary movement, selectable kV/mA level and maximum radiation time 8.1.5.1 S002: Test step 1 ............255 8.1.5.2 S002: Test step 3 ............256 8.1.5.3 S002: Test step 4 ............258 8.1.6...
  • Page 8 Table of contents Sirona Dental Systems GmbH Service Manual ORTHOPHOS SL 8.1.15 S018: Service for height adjustment ..........8.1.15.1 S018: Test step 2............8.1.15.2 S018: Test step 3............8.1.15.3 S018: Test step 4............8.1.15.4 S018: Test step 5............8.1.15.5 S018: Test step 6............
  • Page 9 Sirona Dental Systems GmbH Table of contents Service Manual ORTHOPHOS SL Repair ........................Safety checks ....................Height adjustment motor (M1_4)/spindle............9.2.1 Preparing for motor replacement............. 9.2.1.1 Moving the slide manually ..........9.2.2 Removing board DX32 ..............9.2.3 Replacing the height adjustment motor/spindle ......
  • Page 10 Table of contents Sirona Dental Systems GmbH Service Manual ORTHOPHOS SL 9.7.4 Laser area display (VOL) ..............9.7.4.1 Replacing the laser area display (VOL) ......9.7.4.2 Adjusting the laser area display (VOL) ......9.7.5 Checking the connection of the MS and FH (pan) laser light localizers Occlusal bite block ..................
  • Page 11 Sirona Dental Systems GmbH Table of contents Service Manual ORTHOPHOS SL 9.14 Light barriers ....................393 9.14.1 Replacing the light barriers .............. 393 9.14.2 What has to be done after replacing the light barriers? ....394 9.15 Boards......................395 9.15.1 Important notes about replacing boards........... 395 9.15.2 Replacing boards ................
  • Page 12 Table of contents Sirona Dental Systems GmbH Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images ................. 427 10.11.1 Calling "Test exposures" menu ............428 10.11.2 2D test exposures ................429 10.11.2.1 Pan - diaphragm test exposure (2D) ......429 10.11.2.2 Pan - Symmetrical test exposure (2D)......430 10.11.2.3 Pan - quality test exposure (2D) ........
  • Page 13: About This Service Manual

    ● SIDEXIS 4 Installation Instructions: REF 64 47 200 Current Service Documentation, such as the Service Manual, can be DVD text downloaded from the Sirona dealer website. Scope This Service Manual describes the servicing of the ORTHOPHOS SL digital volume tomograph. It is intended for use exclusively by trained and authorized distributors and service technicians.
  • Page 14: Structure Of The Document

    1 About this Service Manual Sirona Dental Systems GmbH 1.4 Structure of the document Service Manual ORTHOPHOS SL Structure of the document 1.4.1 Identification of the danger levels To prevent personal injury and material damage, please observe the warning and safety information provided in these operating instructions.
  • Page 15: Safety Instructions

    For reasons of product safety, this product may be operated only with original Sirona accessories or third-party accessories expressly approved by Sirona. The user is responsible for any damage resulting from the use of non-approved accessories. Fixed connection...
  • Page 16: Electrostatic Discharge

    2 Safety instructions Sirona Dental Systems GmbH 2.5 Electrostatic discharge Service Manual ORTHOPHOS SL Electrostatic discharge Electrostatic discharge (abbreviated: ESD – ElectroStatic Discharge) Electrostatic discharge from people can damage electronic components when the components are touched. Touch a ground point to discharge static electricity before touching any boards.
  • Page 17: Radiation Protection

    Sirona Dental Systems GmbH 2 Safety instructions Service Manual ORTHOPHOS SL 2.11 Radiation protection 2.11 Radiation protection The valid radiation protection regulations and measures must be Safety information for radiation protection: Service engineer observed. The statutory radiation protection equipment must be used.
  • Page 18: Unit Description

    This symbol can be found next to the rating plate on the unit. Meaning: Observe the Operating Instructions when operating the unit. This symbol can be found on the rating plate on the unit. Meaning: The accompanying documents are available on the Sirona homepage. Electrostatic discharge (ESD) Connector pins or sockets bearing ESD warning labels must not be touched or interconnected without ESD protective measures.
  • Page 19: System Version

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.1 Hardware 3.1.2 System version Version Description Digital X-ray unit with DCS sensor ORTHOPHOS SL 2D Pan sensor: Digital CdTe sensor with direct converter technology for panoramic exposure technology Digital X-ray unit with digital Flat...
  • Page 20: Sensor Versions

    3 Unit description Sirona Dental Systems GmbH 3.1 Hardware Service Manual ORTHOPHOS SL Version Description ORTHOPHOS SL 2D/3D Digital X-ray unit ORTHOPHOS SL Ceph, left-arm version (all variants A) with cephalometer, left-arm version ORTHOPHOS SL 2D/3D Digital X-ray unit ORTHOPHOS SL Ceph, right-arm version...
  • Page 21: Installation Versions

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.1 Hardware 3.1.4 Installation versions Standard installation without remote control with release button on the coiled cable Unit in the treatment room. Installation version 1 with remote control without release Unit...
  • Page 22: Modules And Components

    3 Unit description Sirona Dental Systems GmbH 3.1 Hardware Service Manual ORTHOPHOS SL 3.1.5 Modules and components Slide [ → 23] Stand [ → 24] Cephalometer [ → 25] Remote control [ → 25] 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 23: Slide

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.1 Hardware Slide 3.1.5.1 DX11 M1, M2 DX6* DX83 DX61* DX51 DX78 DX7* DX77 Component Designation Function Boards Open loop/closed loop control in general DX11 Controller board Headrest adapter DX51 Occlusal bite block...
  • Page 24: Stand

    3 Unit description Sirona Dental Systems GmbH 3.1 Hardware Service Manual ORTHOPHOS SL Component Designation Function DX61* Diaphragm control DX77 Control board ambient light DX83 Control board sensor unit Motor Rotary movement of rotating element MAK1, MAK2 Linear movement of rotating element...
  • Page 25: Cephalometer

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.1 Hardware Cephalometer 3.1.5.3 DX91 Component Designation Function Boards DX91 Cephalometer control Motors Stepping motors: linear movement of secondary diaphragm and sensor Light barriers Position check Remote control 3.1.5.4 Component Designation...
  • Page 26: Cabling Overview

    3 Unit description Sirona Dental Systems GmbH 3.1 Hardware Service Manual ORTHOPHOS SL 3.1.6 Cabling overview Netzschalter / S1 Power switch / S1 Interrupteur á bascule / S1 Interruptor de red /S1 DX32 Netzfilter Power Filter Filtre secteur Filtro de red Wagoklemme...
  • Page 27: Sirona Dental Systems Gmbh

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.1 Hardware DX61 LS-Sensor, HV-Motor / Filter Endschalter HV-Motor / Filter oben und unten Moteur HV / Filtre LS-Sensor, Limit switch Motor HV / Filtro above and down Sensor LS, Interrupteur de fin...
  • Page 28: Sirona Dental Systems Gmbh

    3 Unit description Sirona Dental Systems GmbH 3.1 Hardware Service Manual ORTHOPHOS SL DX91 Schrumpfschlauch Shrink hose Gaine thermorétrac. Macarrón termoencogible DX51 Haube, blau Haube, blau Cover, blue Cover, blue Capot, bleu Capot, bleu Tapa, azul Tapa, azul DX1 / J901 DX1 / X500...
  • Page 29: Sirona Dental Systems Gmbh

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.1 Hardware L83 (SUB-D) L83.1 L83.2 DX78 L83.1 L83.2 L83.3 L83.3 DX77 DX83 Sensor Sensor Sensor Sensor Sensor Sensor Sensor Sensor 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 30: Board Photos

    3 Unit description Sirona Dental Systems GmbH 3.1 Hardware Service Manual ORTHOPHOS SL 3.1.7 Board photos Boards in the slide 3.1.7.1 DX1/DX11 boards 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 31: Sirona Dental Systems Gmbh

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.1 Hardware Board DX5 Board DX51 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 32: Sirona Dental Systems Gmbh

    3 Unit description Sirona Dental Systems GmbH 3.1 Hardware Service Manual ORTHOPHOS SL Board DX6 This board is not available as a spare part or a repair part. X-ray tube assemblies can only be ordered as complete units. 64 95 258 D3632...
  • Page 33: Sirona Dental Systems Gmbh

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.1 Hardware DX61 This board is not available as a spare part or a repair part. The diaphragm unit can only be ordered as a complete unit. 64 95 258 D3632...
  • Page 34: Sirona Dental Systems Gmbh

    3 Unit description Sirona Dental Systems GmbH 3.1 Hardware Service Manual ORTHOPHOS SL Board DX7 The board is not available as a spare part or a repair part. The Easypad can only be ordered as a complete unit. Board DX77 64 95 258 D3632...
  • Page 35: Sirona Dental Systems Gmbh

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.1 Hardware Board DX78 Board DX83 / DX831 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 36: Boards In The Stand

    3 Unit description Sirona Dental Systems GmbH 3.1 Hardware Service Manual ORTHOPHOS SL Boards in the stand 3.1.7.2 Board DX32 Line filter 3.1.7.3 Boards in the cephalometer (left- and right-arm versions) Board DX91 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 37: Board In The Remote Control

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.1 Hardware Board in the remote control 3.1.7.4 Board DX42 This board is not available as a spare part or a repair part. 3.1.8 Covers When removing covers, always remember that direct sunlight or bright room lighting can cause system malfunctions due to activated light barriers.
  • Page 38: Sirona Dental Systems Gmbh

    3 Unit description Sirona Dental Systems GmbH 3.1 Hardware Service Manual ORTHOPHOS SL 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 39: Sirona Dental Systems Gmbh

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.1 Hardware Profile covers, top and bottom Intermediate piece Arm cover, top Slide cover, top rear Slide cover, bottom rear Slide cover, center rear Sensor cover, complete Support cover, bottom Support cover, top...
  • Page 40: Technical Data

    3 Unit description Sirona Dental Systems GmbH 3.1 Hardware Service Manual ORTHOPHOS SL 3.1.9 Technical data Unit data 3.1.9.1 Model designation: ORTHOPHOS XG 3D/Ceph Nominal voltage: 200 – 240 V Permissible fluctuation: ± 10% Permissible drop under load: 10% Rated current: Nominal power output:...
  • Page 41: Sirona Dental Systems Gmbh

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.1 Hardware Marking of focal spot: Source-skin distance > 200 mm (8“) Automatic exposure The duration of automatic exposure blocking: blocking (cooling period) depends on the set kV/mA level and the actual exposure time.
  • Page 42: Operating And Transport Conditions

    3 Unit description Sirona Dental Systems GmbH 3.1 Hardware Service Manual ORTHOPHOS SL DCS panorama sensor (optional) Digital CdTe sensor with direct converter technology (DCS) for panoramic exposure technology Active sensor area, Pan type: 146 x 6 mm Pixel size: 0.1 mm...
  • Page 43: Diagrams

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.1 Hardware Diagrams 3.1.9.3 Diagrams, 2D XG 3D Plus Cooling curve for tube housing Cooling curve of X-ray tube SR 90/15 FN Heating curve of tube housing Central X-ray beam and anode angle 7°...
  • Page 44: Requirements For The Pc System

    3 Unit description Sirona Dental Systems GmbH 3.1 Hardware Service Manual ORTHOPHOS SL Requirements for the PC system 3.1.9.4 Requirements on workstation PCs when using a separate RCU Requirements for workstation PCs, service engineer Requirements Requirements Requirements for Fulfilled for 2D workstation...
  • Page 45: Firmware

    Sirona Dental Systems GmbH 3 Unit description Service Manual ORTHOPHOS SL 3.2 Firmware Firmware Any software combinations other than those listed here are not allowed. If a module software version does not match the main software version, the main software version is identified with an asterisk on the info screen (e.g.
  • Page 46: Sirona Dental Systems Gmbh

    3 Unit description Sirona Dental Systems GmbH 3.2 Firmware Service Manual ORTHOPHOS SL SIDEXIS 4 ORTHOPHOS SL GALAXIS GALILEOS workstation software acquisition server software Implant V4.1.2 V1.0.3 V1.0.2 V1.9 SP1 V1.9 SP1 Main software V 05.10.02 ORTHOPHOS SL Cephalometer Remote control Board...
  • Page 47: General Operating Procedures

    Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.1 Switching the unit on General operating procedures Switching the unit on WARNING X-rays Be sure to observe the radiation protection regulations applicable in your country. ➢ No person may be positioned in the unit when it is switched on.
  • Page 48: Factory Setting After Switch-On

    4 General operating procedures Sirona Dental Systems GmbH 4.1 Switching the unit on Service Manual ORTHOPHOS SL IMPORTANT After switch-on, the unit requires approx. 1 min. of warm-up time. At the same time, the self-adjustment routine for the mechanics and the electronics of the unit is performed. If a key is pressed during the self- adjustment routine, an error message will display on the control panel.
  • Page 49: Setting Up The X-Ray Component

    Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.2 Setting up the X-ray component Setting up the X-ray component Since it is addressable via the network, the X-ray component can in principle be activated for X-ray image acquisition by any of the PCs connected to the network.
  • Page 50: Sirona Dental Systems Gmbh

    4 General operating procedures Sirona Dental Systems GmbH 4.2 Setting up the X-ray component Service Manual ORTHOPHOS SL Adding an X-ray component ✔ SIDEXIS 4 is installed. ✔ "Sirona AcquisitionServer" software is installed. ✔ The workstation/plug-in software for the corresponding X-ray component or sensor is installed on the workstation.
  • Page 51: Sirona Dental Systems Gmbh

    Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.2 Setting up the X-ray component Open the menu for adding X-ray components If necessary: "Acquisition Select the acquisition server from the Server" list field. "Next" 5. Click on the button.
  • Page 52: Sirona Dental Systems Gmbh

    4 General operating procedures Sirona Dental Systems GmbH 4.2 Setting up the X-ray component Service Manual ORTHOPHOS SL Automatic search ✔ The menu for adding X-ray components opens. "Automatic search" 1. Click the button in the structure tree. The network is searched for available X-ray components.
  • Page 53: Sirona Dental Systems Gmbh

    Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.2 Setting up the X-ray component Added X-ray components in the window "Device Administration" The X-ray component appears in the device list of the "ORTHOPHOS SL" "Device Administration" configuration menu "Cancel"...
  • Page 54: Sirona Dental Systems Gmbh

    4 General operating procedures Sirona Dental Systems GmbH 4.2 Setting up the X-ray component Service Manual ORTHOPHOS SL Boot service IMPORTANT The X-ray component can only be detected via the boot service if the default network data is set on the unit (see chapter “S037: Test step 2 [ →...
  • Page 55: Sirona Dental Systems Gmbh

    Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.2 Setting up the X-ray component Adapting the network data 3. Adapt the network configuration (IP address, subnet mask and the standard gateway) to your network. "Save" 4. Click the button to save the settings.
  • Page 56: Sirona Dental Systems Gmbh

    4 General operating procedures Sirona Dental Systems GmbH 4.2 Setting up the X-ray component Service Manual ORTHOPHOS SL Added X-ray components in the window "Device Administration" The X-ray component appears in the device list of the "ORTHOPHOS SL" "Device Administration" configuration menu "Cancel"...
  • Page 57: Sirona Dental Systems Gmbh

    Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.2 Setting up the X-ray component Add device manually ✔ The menu for adding X-ray components opens. "Add device manually" "Add device manually" 1. Click the button in the structure tree.
  • Page 58: Sirona Dental Systems Gmbh

    4 General operating procedures Sirona Dental Systems GmbH 4.2 Setting up the X-ray component Service Manual ORTHOPHOS SL Added X-ray components in the window "Device Administration" The X-ray component appears in the device list of the "ORTHOPHOS SL" "Device Administration" configuration menu "Cancel"...
  • Page 59: Opening / Saving "Unit Information

    Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.3 Opening / saving "Unit information" Opening / saving "Unit information" ✔ SIDEXIS 4 is installed. ✔ The workstation software for the corresponding X-ray component or sensor is installed on the workstation.
  • Page 60: Sirona Dental Systems Gmbh

    4 General operating procedures Sirona Dental Systems GmbH 4.3 Opening / saving "Unit information" Service Manual ORTHOPHOS SL "Device configuration" menu opens. "Device details" 4. Click on the button in the structure tree and then on "Show Information" button. 64 95 258 D3632...
  • Page 61: Sirona Dental Systems Gmbh

    Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.3 Opening / saving "Unit information" The Unit information (Advanced information) is displayed. "Downloading…" 5. Click on the button to load and save the current unit information for the unit.
  • Page 62: Activating Functions

    4 General operating procedures Sirona Dental Systems GmbH 4.4 Activating functions Service Manual ORTHOPHOS SL Activating functions 4.4.1 Activation via Easypad Function enhancements If you have purchased a certificate with a corresponding activation key code, you can extend the functionality of your ORTHOPHOS SL system.
  • Page 63: Sirona Dental Systems Gmbh

    Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.4 Activating functions Checking the function enhancement ✔ Level 1 is displayed on the touchscreen and the message “H403 – Switch SIDEXIS to ready for exposure state” is displayed in the comment line.
  • Page 64: Activation Via Pc

    3. Confirm your input with the "Enter" key. A security prompt appears. "Continue to this Website" 4. Click on the link NOTE: Javascript must be enabled. The Sirona Web service is loaded. "User" The sub menus of the menu are displayed in the structure tree.
  • Page 65 Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.4 Activating functions "ActivationCode" 5. In the structure tree, click on the element. A page with an input field is opened. 6. Enter the activation key code from the certificate in the field "ActivationCode"...
  • Page 66 4 General operating procedures Sirona Dental Systems GmbH 4.4 Activating functions Service Manual ORTHOPHOS SL Certificate check The first time you establish contact with the X-ray device, a certificate check (operating system function) should be performed. If no certificate "Security Alert" check has been performed, the dialog box is displayed every time a connection is established to the X-ray device.
  • Page 67 Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.4 Activating functions "Automatically select the certificate store based on the 4. Select the type of certificate" "Next" option panel. Click on the button. "Certificate Import Wizard" "Completing the Certificate...
  • Page 68: Updating The Unit Firmware

    IMPORTANT Read the information provided on the firmware CD supplied with the unit and on the SIRONA dealer page on the Internet very carefully. These sources always contain the latest information on software updates. ✔...
  • Page 69 The update file is located on the unit software CD. It is delivered with each DX11 replacement board and also included in the country set. The contents of the CD can be downloaded from the Dealer domain of the SIRONA Internet home page (under Products/Imaging systems): www.sirona.com "Update starten"...
  • Page 70 “ProgramData ⇒ Sirona ⇒ AcquisitionServer ⇒ SWLog” 8. Check the log file to make sure that the update was completed successfully. If the log file contains entries such as "Update of DXxx failed!", please repeat the update.
  • Page 71: Backing Up / Restoring / Installing Sensor Data

    Service Manual ORTHOPHOS SL 4.6 Backing up / restoring / installing sensor data Backing up / restoring / installing sensor data The sensor data can be saved, resaved and installed using the Sirona Web service. 4.6.1 Access the “Backup / restore sensor data” menu ✔...
  • Page 72 4 General operating procedures Sirona Dental Systems GmbH 4.6 Backing up / restoring / installing sensor data Service Manual ORTHOPHOS SL The dialog box for entering the access data opens. 6. Enter the following access data: Username: “service” Password: "sirona" "OK" 7. Confirm your entry by clicking the button.
  • Page 73: Backing Up Sensor Data

    Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.6 Backing up / restoring / installing sensor data 4.6.2 Backing up sensor data "Backup sensor data" You can use the menu to load the sensor data Backing up sensor data1 (factory calibration of the sensor) from the DX83 board and save it as a Zip file.
  • Page 74: Restoring Sensor Data

    4 General operating procedures Sirona Dental Systems GmbH 4.6 Backing up / restoring / installing sensor data Service Manual ORTHOPHOS SL 4.6.3 Restoring sensor data Previously saved sensor data (factory calibration of the sensor) can be "Restore sensor data" restored on the DX83 board using the menu.
  • Page 75: Installing Sensor Data

    Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.6 Backing up / restoring / installing sensor data 4.6.4 Installing sensor data Sensor data (factory calibration of the sensor) of a new sensor can be "Install new sensor data"...
  • Page 76: Calling The Info Screen

    4 General operating procedures Sirona Dental Systems GmbH 4.7 Calling the Info screen Service Manual ORTHOPHOS SL Calling the Info screen The Info screen lists device data that is useful for any discussions with your service engineer. ✔ The touchscreen display is located on level 1.
  • Page 77: Using Demo Mode - Operation Without Radiation Release

    Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.8 Using demo mode – operation without radiation release Using demo mode – operation without radiation release If the unit is to be presented as a demo unit at trade fairs or exhibitions, it must be ensured that radiation release is blocked.
  • Page 78: Switching Off Demo Mode

    4 General operating procedures Sirona Dental Systems GmbH 4.8 Using demo mode – operation without radiation release Service Manual ORTHOPHOS SL 4.8.2 Switching off demo mode 1. Switch off the unit. DANGER Perilous shock hazard! It is essential to switch off the unit and to wait at least another 4 minutes before taking off the covers of the X-ray tube assembly.
  • Page 79: Important Information For Repacking And Transport

    Sirona Dental Systems GmbH 4 General operating procedures Service Manual ORTHOPHOS SL 4.8 Using demo mode – operation without radiation release 4.8.3 Important information for repacking and transport IMPORTANT If a used carton on which one of the shockwatch or tiltwatch indicators has already been tripped is used to package the unit, please make an entry to that effect on the delivery note.
  • Page 80: Messages

    5 Messages Sirona Dental Systems GmbH 5.1 Help messages Service Manual ORTHOPHOS SL Messages The different message texts are displayed ... ● on the touchscreen of the Easypad ● On the display of the remote control There are 3 groups of message texts: Help messages (Hx xx): ●...
  • Page 81: Status Displays

    Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.2 Status displays Help code Description Actions required "Switch SIDEXIS to ready for exposure state" H4 03 ● Establish readiness for exposure in SIDEXIS "Plug in Ceph sensor" H4 04 ● Plug the ceph sensor into the cephalometer.
  • Page 82: Error Messages

    5 Messages Sirona Dental Systems GmbH 5.3 Error messages Service Manual ORTHOPHOS SL Error messages Error messages are displayed as error codes (Ex yy zz) on the Easypad touchscreen and on the remote control display (if there is one). The codes provide you with error type, error location and troubleshooting information.
  • Page 83: Yy - Location

    Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.3 Error messages Description Error group Actions required Malfunction or This error group includes all errors ● Acknowledge the error message and mechanical obstruction that indicate problems with the make sure that the movements of the unit...
  • Page 84: List Of Error Messages

    Description Actions required E6 06 01 E6 06 01 General error during module ● Please contact the Sirona Customer S. [ → 68] initialization Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 85 Module failed in TTP (detected on ● Check the CAN bus. S. [ → 141] master side) ● Please contact the Sirona Customer S. [ → 68] This error may also occur in Service Center (CSC) to find out connection with other causal error...
  • Page 86 5 Messages Sirona Dental Systems GmbH 5.4 List of error messages Service Manual ORTHOPHOS SL Error code Description Actions required E7 06 12 E7 06 12 Unit is not ready for operation ● Check the CAN bus. S. [ → 141] If this error occurs in combination with This error may also occur in other errors ...
  • Page 87 If a CAN bus error had been reported before ... ● Check the CAN bus. S. [ → 141] ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 88 5 Messages Sirona Dental Systems GmbH 5.4 List of error messages Service Manual ORTHOPHOS SL Error code Description Actions required E1 06 41 E1 06 41 Tolerance exceeded kV nom. ● Run service routine S005.2. S. [ → 260] ● Replace the tube assembly.
  • Page 89 Actions required E6 06 56 E6 06 56 Error during auto-compensation. ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 90: Location 07: Easypad/Dx7

    Description Actions required E6 07 01 E6 07 01 General error during module ● Please contact the Sirona Customer S. [ → 68] initialization Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 91 Module failed in TTP (detected on ● Check the CAN bus. S. [ → 141] master side). ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out This error may also occur in whether a bug fix by means of a...
  • Page 92 5 Messages Sirona Dental Systems GmbH 5.4 List of error messages Service Manual ORTHOPHOS SL Error code Description Actions required E7 07 12 E7 07 12 Unit is not ready for operation ● Check the CAN bus. S. [ → 141] This error is a sequential fault.
  • Page 93: Location 10: System Hardware

    Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.4 List of error messages 5.4.3 Location 10: System hardware Error code Description Actions required E7 10 01 E7 10 01 EEPROM cannot be written. ● Acknowledge error and repeat procedure. If the error occurs repeatedly ...
  • Page 94: Location 11: Power Pc/Board Dx11

    Actions required E6 11 01 E6 11 01 Program sequence error. ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 95 ● If necessary, obtain a new remote control functions. To continue to work with from the manufacturer. this unit, you must unplug the A remote control for another Sirona unit or a remote control and restart the unit. third-party manufacturer unit may have been connected.
  • Page 96: Location 12: Can Bus

    5 Messages Sirona Dental Systems GmbH 5.4 List of error messages Service Manual ORTHOPHOS SL Error code Description Actions required E1 11 20 E1 11 20 The calibration data on the unit is ● Perform unit adjustment and calibration. S. [ → 160] invalid or does not match the serial ●...
  • Page 97: Location 13: Stand/Peripherals

    Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.4 List of error messages 5.4.6 Location 13: Stand/Peripherals Error code Description Actions required E4 13 01 E4 13 01 Actuator 1 has not reached pan ● Check the actuator mechanisms manually home position.
  • Page 98 1. Switch off the unit. 2. Wait 1 minute. 3. Switch on the unit. 4. Repeat procedure. ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 99 Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.4 List of error messages Error code Description Actions required E4 13 11 E4 13 11 Actuator 2 has not reached pan ● Check the actuator mechanisms manually home position. for smooth and easy running.
  • Page 100 1. Switch off the unit. 2. Wait 1 minute. 3. Switch on the unit. 4. Repeat procedure. ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 101 Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.4 List of error messages Error code Description Actions required E4 13 22 E4 13 22 Ring motor has not left pan home ● Check the ring drive mechanism manually S. [ → 414], position.
  • Page 102 3. Switch on the unit. 4. Repeat procedure and observe causal error messages. ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 103 Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.4 List of error messages Error code Description Actions required E4 13 30 E4 13 30 No height adjustment motor pulses. ● Check cable L16 (X402), replace if S. [ → 153], necessary.
  • Page 104 2. Wait 1 minute. 3. Switch unit on. 4. Repeat procedure and check function. ● Please contact the Sirona Customer S. [ → 45] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 105 Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.4 List of error messages Error code Description Actions required E3 13 40 E3 13 40 Release signal applied during ● Restart the unit: power-on. 1. Switch off the unit. 2. Wait 1 minute.
  • Page 106 2. Wait 1 minute. 3. Switch on the unit. 4. Repeat procedure and check function. ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 107 Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.4 List of error messages Error code Description Actions required E5 13 60 E5 13 60 Temple supports; have not left ● Check light barriers V5_1 to V5_4, replace S. [ → 151], home position despite movement headrest and light barrier if necessary.
  • Page 108 2. Wait 1 minute. 3. Switch on the unit. 4. Repeat procedure and check function. ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 109 2. Wait 1 minute. 3. Switch on the unit. 4. Repeat procedure and check function. ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 110: Location 14: Digital Extension, Sidexis 4

    ● Send in Xab.ini and raw image of last S. [ → 267], from SIDEXIS 4. exposure to the Sirona Customer Service S. [ → 68] Center (CSC) (check the binning setting) and seek advice on how to proceed from the CSC.
  • Page 111 Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.4 List of error messages Error code Description Actions required E6 14 06 E6 14 06 The bootline of board DX11 had to ● Re-configuration of network data required. be preassigned with default values.
  • Page 112: Location 15: Configuration, Update

    Description Actions required E6 15 10 E6 15 10 Update file for module is corrupt. ● Obtain latest update file from the Sirona S. [ → 68] Customer Service Center (CSC) or the Sirona home page and perform software update.
  • Page 113: Location 20: Sensor Unit, General

    4. Repeat procedure. If the error occurs repeatedly ... S. [ → 68] ● Please contact the Sirona Customer Service Center (CSC) to find out whether a bug fix by means of a firmware update is possible and perform such an update if necessary.
  • Page 114 S. [ → 59] entries in the "DX831" area. If the error occurs repeatedly ... ● Please contact the Sirona Customer Service Center (CSC) to find out whether a bug fix by means of a firmware update is possible and perform such an update if necessary.
  • Page 115 (KSC). If the error occurs repeatedly ... S. [ → 68] ● Please contact the Sirona Customer Service Center (CSC) to find out whether a bug fix by means of a firmware update is possible and perform such an update if necessary.
  • Page 116 "DX831" area. If the error occurs repeatedly ... S. [ → 68] ● Please contact the Sirona Customer Service Center (CSC) to find out whether a bug fix by means of a firmware update is possible and perform such an update if necessary.
  • Page 117 Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.4 List of error messages Error code Description Actions required E7 20 41 E7 20 41 Several fans in the sensor unit are ● Check fans in the sensor unit or plug S.
  • Page 118: Location 22: Dcs Sensor

    5 Messages Sirona Dental Systems GmbH 5.4 List of error messages Service Manual ORTHOPHOS SL 5.4.10 Location 22: DCS sensor Error code Description Actions required E6 22 01 E6 22 01 DCS sensor could not be initialized. ● Check correct positioning of the DCS sensor on board DX83.
  • Page 119: Location 23: Flat Panel Detector

    Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.4 List of error messages 5.4.11 Location 23: Flat Panel Detector Error code Description Actions required E6 23 01 E6 23 01 Error in initializing the digital flat ● Check the connection of the flat panel panel detector.
  • Page 120: Location 42: Remote Control

    Description Actions required E6 42 01 E6 42 01 General module initialization error. ● Please contact the Sirona Customer S. [ → 68], Service Center (CSC) to find out whether Error generated during module self- S. [ → 395] a bug fix by means of a software update is test.
  • Page 121 DX42, replace module if necessary S. [ → 395] ● Replace board DX42. ● Please contact the Sirona Customer Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 122 5 Messages Sirona Dental Systems GmbH 5.4 List of error messages Service Manual ORTHOPHOS SL Error code Description Actions required E7 42 12 E7 42 12 Unit is not ready for operation This error is a sequential fault. ● Restart the unit: 1.
  • Page 123 Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.4 List of error messages Error code Description Actions required E3 42 30 E3 42 30 R key pressed during power-on. ● Restart the unit: 1. Switch off the unit. 2. Wait 1 minute.
  • Page 124: Location 61: Diaphragm Control, Board Dx61

    Description Actions required E6 61 01 E6 61 01 General error during module ● Please contact the Sirona Customer S. [ → 68], initialization Service Center (CSC) to find out whether S. [ → 146], a bug fix by means of a software update is possible and perform such an update if S.
  • Page 125 TTP = Time Trigger Protocol diaphragm unit, if necessary. S. [ → 370], ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 126 5 Messages Sirona Dental Systems GmbH 5.4 List of error messages Service Manual ORTHOPHOS SL Error code Description Actions required E7 61 12 E7 61 12 Unit is not ready for operation This error is a sequential fault. S. [ → 141], ●...
  • Page 127 S. [ → 393], ● Perform diaphragm adjustment. S. [ → 160], ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 128 S. [ → 151], ● Perform diaphragm adjustment. S. [ → 160], ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 129 Sirona Dental Systems GmbH 5 Messages Service Manual ORTHOPHOS SL 5.4 List of error messages Error code Description Actions required E5 61 55 E5 61 55 Rotating element position not ● Check light barriers V61_4; replace if S. [ → 151], reached.
  • Page 130: Location 81, Board Dx81, Ceph Slot

    Actions required E6 81 51 E6 81 51 Ceph: General error during module ● Please contact the Sirona Customer S. [ → 68] initialization Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 131 ● Check ceph sensor (replacement sensor), replace if necessary. ● Please contact the Sirona Customer Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 132 Ceph: The sensor type on the ceph ● Check the SIDEXIS 4 specifications with S. [ → 68] slot is not compatible with SIDEXIS the Sirona Customer Service Center 4 specifications. (CSC). ● Check the version compatibility of the unit software and SIDEXIS 4, perform software update if necessary.
  • Page 133 ● Run software update. ● Check the version compatibility of the unit software and SIDEXIS 4, perform software update if necessary. ● Seek advice from the Sirona Customer Service Center (CSC) with regard to the settings/sensor information in the Xab.ini file.
  • Page 134: Location 83: Sensor Unit

    5 Messages Sirona Dental Systems GmbH 5.4 List of error messages Service Manual ORTHOPHOS SL 5.4.15 Location 83: Sensor unit Error code Description Actions required E7 83 01 E7 83 01 Malfunction when starting the ● Check plug connections of fiber-optic S. [ → 30] sensor unit.
  • Page 135: Location 91: Cephalometer, Board Dx91

    Description Actions required E6 91 01 E6 91 01 General error during module ● Please contact the Sirona Customer S. [ → 68], initialization Service Center (CSC) to find out whether S. [ → 396] a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 136 ● Replace board DX91. S. [ → 396], TTP = Time Trigger Protocol ● Please contact the Sirona Customer S. [ → 68] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 137 S. [ → 404], ● Check board DX91, replace if necessary. S. [ → 146], ● Please contact the Sirona Customer S. [ → 396] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 138 S. [ → 404], ● Check board DX91, replace if necessary. S. [ → 146], ● Please contact the Sirona Customer S. [ → 396] Service Center (CSC) to find out whether a bug fix by means of a software update is possible and perform such an update if necessary.
  • Page 139: Troubleshooting

    Sirona Dental Systems GmbH 6 Troubleshooting Service Manual ORTHOPHOS SL 6.1 Error logging memory Troubleshooting DANGER Potentially lethal shock hazard! It is essential to switch off the unit and wait at least 1 minute before removing a cover. Switch OFF the X-ray unit before connecting a measuring instrument.
  • Page 140: Example Of Error Logging Data

    ● Rescue request Sidexis Timeout These entry data may also occur after "Recording stopped" or "Cancel" and indicate an exceptional circumstance. You can supply important information for error diagnosis in coordination with the Sirona Customer Service Center. Entry data [Error] E6 07 06...
  • Page 141: Checking The Can Bus

    Detailed error bytes for an error 0x10 0x00 0x00 0x00 0x00 0x00 0x00 occurrence 0x00 Entry data SidErr: ERR_SOCKET_ERROR Detail of network error (for Sirona only) [Error Sidexis] SockErr: Detail of network error (for Sirona only) Entry data [Stringname] Key Act...
  • Page 142 6 Troubleshooting Sirona Dental Systems GmbH 6.2 Checking the CAN bus Service Manual ORTHOPHOS SL 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 143 Sirona Dental Systems GmbH 6 Troubleshooting Service Manual ORTHOPHOS SL 6.2 Checking the CAN bus 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 144 6 Troubleshooting Sirona Dental Systems GmbH 6.2 Checking the CAN bus Service Manual ORTHOPHOS SL 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 145: Checking The Can Bus With The Diagnostic Function Of Board Dx1

    Sirona Dental Systems GmbH 6 Troubleshooting Service Manual ORTHOPHOS SL 6.2 Checking the CAN bus 6.2.1 Checking the CAN bus with the diagnostic function of board DX1 Board DX1 features a diagnostic function for diagnosing malfunctions of the CAN bus via LEDs V700 and V701 (see wiring diagrams). The...
  • Page 146: Checking The Boards

    6 Troubleshooting Sirona Dental Systems GmbH 6.3 Checking the boards Service Manual ORTHOPHOS SL Checking the boards 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 147 Sirona Dental Systems GmbH 6 Troubleshooting Service Manual ORTHOPHOS SL 6.3 Checking the boards Important LEDs on the boards (see also wiring diagrams) Board LEDs Normal mode Malfunction Bootloader V200 flashing at 1 Hz lit / not lit flashing quickly 64 90 960...
  • Page 148: Checking Board Dx32

    6 Troubleshooting Sirona Dental Systems GmbH 6.3 Checking the boards Service Manual ORTHOPHOS SL 6.3.1 Checking board DX32 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 149 Sirona Dental Systems GmbH 6 Troubleshooting Service Manual ORTHOPHOS SL 6.3 Checking the boards 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 150: Checking The Motors

    6 Troubleshooting Sirona Dental Systems GmbH 6.4 Checking the motors Service Manual ORTHOPHOS SL Checking the motors 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 151: Checking The Light Barriers

    Sirona Dental Systems GmbH 6 Troubleshooting Service Manual ORTHOPHOS SL 6.5 Checking the light barriers Checking the light barriers 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 152: Device Leakage Current Too High

    6 Troubleshooting Sirona Dental Systems GmbH 6.6 Device leakage current too high Service Manual ORTHOPHOS SL Device leakage current too high 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 153: Checking The Cables

    Sirona Dental Systems GmbH 6 Troubleshooting Service Manual ORTHOPHOS SL 6.7 Checking the cables Checking the cables NOTICE You can use a standard Cat5 cable as a test cable for L10, L12, L40 and L37. This cable must not be permanently installed.
  • Page 154: Checking Fiber-Optic Cables

    6 Troubleshooting Sirona Dental Systems GmbH 6.8 Checking fiber-optic cables Service Manual ORTHOPHOS SL Checking fiber-optic cables ® A fiber test device, e.g. Fiberpoint 250MD is required [ → 13] to test the fiber-optic cables. Checking fiber-optic cables L83B (1) between board DX83 (X700)
  • Page 155 Sirona Dental Systems GmbH 6 Troubleshooting Service Manual ORTHOPHOS SL 6.8 Checking fiber-optic cables Checking fiber-optic cables L83B (2) between board DX83 (X800) and board DX1 (J901) 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 156: Error Analysis Of X-Ray Control Signal Path

    6 Troubleshooting Sirona Dental Systems GmbH 6.9 Error analysis of X-RAY control signal path Service Manual ORTHOPHOS SL Error analysis of X-RAY control signal path with remote control installed Error and help messages E3 42 31 + E3 13 40 occur in...
  • Page 157 Sirona Dental Systems GmbH 6 Troubleshooting Service Manual ORTHOPHOS SL 6.9 Error analysis of X-RAY control signal path E6 13 43 occurs once during operation of the unit: H321 is triggered at start of exposure: 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 158 6 Troubleshooting Sirona Dental Systems GmbH 6.9 Error analysis of X-RAY control signal path Service Manual ORTHOPHOS SL without installed remote control Error messages E3 13 40 occurs after the unit is switched with and without installed remote control Error messages Error code...
  • Page 159: Checking The Data Paths

    Sirona Dental Systems GmbH 6 Troubleshooting Service Manual ORTHOPHOS SL 6.10 Checking the data paths 6.10 Checking the data paths Checking the 2D and 3D data paths is possible through the service routines S032.10 (DCS sensor) and S032.50 (flat panel detector) (see chapter “S032: Sensor test [ →...
  • Page 160: Unit Adjustment And Calibration

    7 Unit adjustment and calibration Sirona Dental Systems GmbH Service Manual ORTHOPHOS SL Unit adjustment and calibration DANGER X-rays When performing the following tests, be sure to observe the radiation protection regulations applicable in your country (see Operating Instructions). DANGER X-rays "Radiation"...
  • Page 161 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 162: General Information On Unit Adjustment And Calibration

    7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.1 General information on unit adjustment and calibration Service Manual ORTHOPHOS SL General information on unit adjustment and calibration Please adhere to the following order when adjusting and calibrating the system: 2D adjustment (panoramic) "Sensor adjustment"...
  • Page 163: Displays And Help Messages During Adjustment And Calibration

    Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.1 General information on unit adjustment and calibration 7.1.1 Displays and help messages during adjustment and calibration The most frequent help and status messages during calibration are listed below.
  • Page 164: Calibration Menu

    7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.1 General information on unit adjustment and calibration Service Manual ORTHOPHOS SL 7.1.2 Calibration menu The menu guides you through the procedure to adjust and calibrate the unit. 7.1.2.1 Opening the calibration menu ✔...
  • Page 165 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.1 General information on unit adjustment and calibration Password protection The calibration menu is password-protected. For the password, enter the first four digits of the current system date (PC) in reverse order. Example: On 25.03.2016, the service password is 3052.
  • Page 166: Menu Structure

    7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.1 General information on unit adjustment and calibration Service Manual ORTHOPHOS SL Menu structure 7.1.2.2 The menu is divided into different functional areas. Navigation area Structure tree for adjustment and calibration [ → 169] Action area Data entry and display window.
  • Page 167 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.1 General information on unit adjustment and calibration Scroll bar Scrolls within the exposure window Message window Shows messages and information about this stage of the adjustment/calibration procedure In addition to the different areas, the menu also contains the following buttons: "Fine adjustment"...
  • Page 168 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.1 General information on unit adjustment and calibration Service Manual ORTHOPHOS SL Toolbar functions Button Function Rotate exposure 90° clockwise. Reducing size of section Increasing size of section Adjusting the brightness Adjusting the contrast...
  • Page 169 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.1 General information on unit adjustment and calibration 7.1.2.2.1 Navigation area The navigation area contains a structure tree similar to the one you will be familiar with from your Windows interface. The structure tree contains all stages of the adjustment and calibration procedure you need to complete in order to adjust and calibrate your system.
  • Page 170 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.1 General information on unit adjustment and calibration Service Manual ORTHOPHOS SL Validity of existing adjustment and calibration The elements of the structure tree used for adjusting (calibrating) the unit are prefixed by status indicators indicating the current status of the corresponding adjustment or calibration operation.
  • Page 171 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.1 General information on unit adjustment and calibration Examples of image acquisition with precision and coarse adjustment "Sensor adjustment" Sensor adjustment can usually be performed directly via precision (Pan - adjustment) adjustment.
  • Page 172 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.1 General information on unit adjustment and calibration Service Manual ORTHOPHOS SL "Diaphragm adjustment" Diaphragm adjustment can usually be performed directly via precision (Pan - adjustment) adjustment. Only in exceptional cases, e.g. if the exposed image area is "Fine...
  • Page 173 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.1 General information on unit adjustment and calibration "Adjustment of Ceph secondary In most cases the ceph secondary diaphragm can be performed directly diaphragm" via precision adjustment. Only in exceptional cases, e.g. if the exposed...
  • Page 174 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.1 General information on unit adjustment and calibration Service Manual ORTHOPHOS SL 7.1.2.2.3 Shifting direction of the exposed image area The menus contain a pictographic representation of the expected adjustment image to help you perform software-based adjustment and calibration.
  • Page 175: Enabling Exposure Readiness

    Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.1 General information on unit adjustment and calibration 7.1.3 Enabling exposure readiness To take an exposure, the system must first be made ready for exposure. ✔ The calibration menu must be opened [ → 164].
  • Page 176: Test Phantom For Adjustment And Calibration

    7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.1 General information on unit adjustment and calibration Service Manual ORTHOPHOS SL 7.1.6 Test phantom for adjustment and calibration Needle phantom for panoramic adjustment 7.1.6.1 NOTICE XGPF-36318; Use needle phantom correctly Risk of damage to unit...
  • Page 177 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.1 General information on unit adjustment and calibration When fitting the needle phantom, make sure that it is correctly oriented. For the panorama adjustment of the unit, the phantom must be fitted so that the needles (C) point upward.
  • Page 178: Adjustment Phantom For Adjustment Of The Cephalometer

    7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.1 General information on unit adjustment and calibration Service Manual ORTHOPHOS SL Adjustment phantom for adjustment of the cephalometer 7.1.6.2 "Adjustment of Ceph primary diaphragm" "Adjustment of Ceph main beam direction" ceph adjustments, the ceph adjustment phantom (A) must be inserted into the sensor unit cover.
  • Page 179: Geometry Phantom For Volume Calibration

    Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.1 General information on unit adjustment and calibration Geometry phantom for volume calibration 7.1.6.3 NOTICE Risk of damage to unit Make absolutely sure that the geometry phantom is inserted in the bite...
  • Page 180 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.1 General information on unit adjustment and calibration Service Manual ORTHOPHOS SL NOTICE For correct calibration, the geometry phantom must be aligned horizontally and vertically using a spirit level after insertion in the bite block holder.
  • Page 181: Adjustment And Calibration Via The Calibration Menu

    After every adjustment and calibration of the unit, the reference values for the constancy measurement must be recalculated and entered in the "Test results" form, "Reference value" column (see technical document "ORTHOPHOS SL 2D/3D Acceptance test DIN 6868 Part 151 / Part 161" (REF 65 61 059).
  • Page 182 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 7.2.1.1.1 Automatic adjustment: Panoramic sensor (standard) ✔ The needle phantom is in the bite block holder on the unit [ → 176].
  • Page 183 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu Adjustment OK The three needle images must lie in the center and inside the auxiliary lines. Adjustment not OK 5. Repeat the procedure starting at Step 3 until the values in the menu's text boxes are equal to "zero".
  • Page 184 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 6. If adjustment is OK (the adjustment values in the text boxes are equal to zero), save the values. [ → 175] "Sensor adjustment"...
  • Page 185 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 7.2.1.1.2 Manual adjustment: Panoramic sensor 1. Start the adjustment procedure as described in the chapter entitled "Automatic adjustment: Panoramic sensor (standard) [ → 182]".
  • Page 186 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 3. Overwrite the adjustment values for S1, S2 and S3 in the text boxes "Sensor adjustment" of the menu with the measured values.
  • Page 187: Pan Aperture Adjustment

    Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu Pan aperture adjustment 7.2.1.2 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 188 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 7.2.1.2.1 Automatic adjustment: Pan diaphragm (standard) IMPORTANT If a message window indicating that the diaphragm is tilted appears in "Diaphragm adjustment"...
  • Page 189 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 190 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL Adjustment OK ● The exposed diaphragm area must lie horizontally centered in the image section as well as inside the auxiliary lines.
  • Page 191 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 7.2.1.2.2 Manual adjustment: Pan - Diaphragm The manual adjustment procedure is similar to the one for automatic adjustment. The only difference is that the default adjustment values "Diaphragm adjustment"...
  • Page 192 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 2. Once you have taken the exposure, measure distances Sx and Sy with the toolbar measuring ruler [ → 166]. Tip: For better measurement, you can also edit [ → 166] the exposure using the toolbar functions.
  • Page 193: Pan Symmetry Adjustment

    Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu Pan symmetry adjustment 7.2.1.3 7.2.1.3.1 Automatic adjustment: Pan - Symmetry ✔ The needle phantom is in the bite block holder on the unit [ → 176].
  • Page 194 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 195 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu Adjustment OK ● The shadow of the center needle, the needle image, and the auxiliary lines must be coincident and located one behind the other.
  • Page 196 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 7.2.1.3.2 Manual adjustment: Pan - Symmetry The manual adjustment procedure is similar to the one for automatic adjustment [ → 182]. The only difference is that the default adjustment "Symmetrie adjustment"...
  • Page 197: Pan Sensor Calibration (Dcs)

    Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu Pan sensor calibration (DCS) 7.2.1.4 NOTE: For sensor calibration, 7 exposures are taken (for the different kV/ mA combinations). The exposure is only displayed in the exposure window at the end of the calibration procedure.
  • Page 198 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 6. Repeat the procedure from point 5 for all other kV/mA combinations. After the last exposure has been taken, a message in the action window indicates whether calibration has been successful and "Sensor calibration...
  • Page 199: Adjustment Of The Ceph Primary Diaphragm

    Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu Adjustment of the ceph primary diaphragm 7.2.1.5 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 200 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 7.2.1.5.1 Automatic adjustment: Ceph - Primary diaphragm (standard) NOTICE Phantom for Ceph exposure Risk of damage to unit It is essential that the needle phantom is removed from the bite block holder on the unit before a ceph exposure is taken;...
  • Page 201 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu S3 = ca. 60mm 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 202 ● Distance S3 must be approx. 60 Adjustment not OK IMPORTANT: If S3is >70mm, please contact the Sirona Customer Service Center: 4. Repeat the procedure starting at Step 2 until the values in the menu's text boxes are equal to "zero". This means that adjustment was performed successfully.
  • Page 203 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 7.2.1.5.2 Manual adjustment: Ceph - Primary diaphragm The manual adjustment procedure is similar to the one for automatic adjustment [ → 200]. The only difference is that the default adjustment "Adjustment of Ceph primary...
  • Page 204: Adjustment Of The Ceph Secondary Diaphragm

    7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL Adjustment of the ceph secondary diaphragm 7.2.1.6 IMPORTANT If you cannot correctly adjust the ceph secondary diaphragm via the calibration menu, the Ceph secondary diagram must first be readjusted mechanically [ →...
  • Page 205 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 7.2.1.6.1 Automatic adjustment: Ceph secondary diaphragm (standard) NOTICE Phantom for Ceph exposure Risk of damage to unit It is essential that the needle phantom is removed from the bite block holder on the unit before a ceph exposure is taken;...
  • Page 206 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL Adjustment OK ● The exposed diaphragm area must lie centered and straight in the image section as well as inside the superimposed auxiliary lines.
  • Page 207 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 4. Repeat the procedure starting at Step 2 until the values in the menu's text boxes are equal to "zero". This means that adjustment was performed successfully.
  • Page 208 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 7.2.1.6.2 Manual adjustment: Ceph secondary diaphragm The manual adjustment procedure is similar to the one for automatic adjustment [ → 205]. The only difference is that the default adjustment "Adjustment of Ceph secondary...
  • Page 209 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 2. Once you have taken the exposure, measure distances S1, S2 and S3 with the toolbar measuring ruler [ → 166].
  • Page 210 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 7.2.1.6.3 Mechanical adjustment: Ceph secondary diaphragm 1. Loosen screw (A) and remove the cover of the secondary diaphragm by pulling it downward.
  • Page 211 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 3. Set the diaphragm tilt via screw (C) (Lx in mm) and the diaphragm height via screw (D) (Ly in mm).
  • Page 212: Adjustment Of The Ceph Secondary Diaphragm

    7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL Adjustment of the ceph secondary diaphragm (QuickShot) 7.2.1.7 7.2.1.7.1 Automatic adjustment: Ceph secondary diaphragm (QuickShot) (standard) NOTICE Phantom for Ceph exposure...
  • Page 213 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 3. Create an exposure (80kV / 14 mA; 0.3 s) [ → 175]. Adjustment value S1 is calculated automatically from the "Adjustment of Ceph...
  • Page 214 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL Adjustment OK ● The exposed image section must lie centered and straight in the image section as well as inside the superimposed auxiliary lines.
  • Page 215 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 216 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 2. Once you have taken the exposure, measure distance S1 with the toolbar measuring ruler [ → 166]. 3. Overwrite the adjustment value for S1 in the text box of the "Adjustment of Ceph secondary diaphragm (Quickshot)"...
  • Page 217: Adjustment Of The Ceph Main X-Ray Beam

    Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu Adjustment of the ceph main X-ray beam 7.2.1.8 7.2.1.8.1 Automatic adjustment: Ceph main X-ray beam direction (standard) NOTICE Risk of damage to unit It is essential that the needle phantom is removed from the pan bite block holder on the unit before a ceph exposure is taken;...
  • Page 218 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL Adjustment value S1 is calculated automatically from the "Adjustment of Ceph exposure and entered in the text box of the main beam direction"...
  • Page 219 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu The exposure is displayed in the exposure window. 4. Repeat the procedure starting at Step 2 until the value in the menu's text box is equal to "zero".
  • Page 220 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 7.2.1.8.2 Manual adjustment: Ceph - Main X-ray beam direction The manual adjustment procedure is similar to the one for automatic adjustment [ → 217]. The only difference is that the default adjustment "Adjustment of Ceph main beam...
  • Page 221: Adjusting The Earplug Alignment

    Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu Adjusting the earplug alignment 7.2.1.9 7.2.1.9.1 Checking the ear plug alignment ✔ The ceph sensor is plugged into the ceph slot on the unit.
  • Page 222 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 3. Fit the adjusting caps (A+B) to the ear plugs and secure them with adhesive tape. Black adjusting cap: on the outside (sensor side) Transparent adjusting cap (B): on the inside (tube assembly side) 4.
  • Page 223 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 7.2.1.9.2 Adjusting the ear plug alignment ➢ Unscrew and remove the cover from the cephalometer. 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 224 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL Horizontal correction 1. Loosen the screws (A) slightly (approx. 2 - 3 turns). Tip: Do not unscrew them fully! 2. Adjust the ear plugs in the horizontal direction by turning the screw (B) counterclockwise or clockwise.
  • Page 225 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 3. Adjust the ear plugs in the vertical direction by turning the knurled nut (F). The direction of rotation for left-arm (G) and right-arm (H) versions is shown in the drawing below.
  • Page 226: 3D Adjustment/Calibration

    7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL 7.2.2 3D adjustment/calibration IMPORTANT For the 3D adjustment and calibration, the values in the input fields do not have to be equal to zero.
  • Page 227 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 2. Align the calibration block with the spirit level by adjusting the screw (A). * For greater clarity, the unit in the left image is displayed without sensor unit.
  • Page 228: Diaphragm Adjustment

    7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL Diaphragm adjustment 7.2.2.2 ✔ is not The 3D calibration block in the bite block holder on the unit. "3D" "Diaphragm 1. In the structure tree, under , click on the adjustment"...
  • Page 229: Sensor Calibration

    Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu Sensor calibration 7.2.2.3 "3D" "Sensor calibration" 1. In the structure tree, under , click on the element (S010.64). 2. Establish [ → 175] receptivity.
  • Page 230: Diaphragm Configuration

    7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL Diaphragm configuration 7.2.2.4 ✔ is not The 3D calibration block in the bite block holder on the unit. "3D" "Diaphragm calibration"...
  • Page 231 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 6. Take an exposure (85 kV / 0.34 mA s) [ → 175]. A message is displayed in the action window to indicate whether adjustment has been successful and the status indicator in front "8x8"...
  • Page 232: Geometry Calibration

    7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL Geometry calibration 7.2.2.5 1. Insert the 3D calibration block into the bite block holder on the unit. [ → 179] 2. Align the 3D calibration block with the spirit level by adjusting the screw (A).
  • Page 233 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 8. Establish [ → 175] receptivity. 9. Take an exposure (85 kV/22 mA s) [ → 175]. A message is displayed in the action window to indicate whether adjustment has been successful and the status indicator in front "5x5"...
  • Page 234: Dosimetry

    7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.2 Adjustment and calibration via the calibration menu Service Manual ORTHOPHOS SL Dosimetry 7.2.2.6 A dosimeter for pulsed radiation (e.g. Mult-O-Meter 512L) is required for dosimetry. "3D" "Dose measurement" 1. In the structure tree, under , click on the element (S011.21).
  • Page 235: Resetting Adjustment/Calibration

    Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.2 Adjustment and calibration via the calibration menu 7.2.3 Resetting adjustment/calibration In an absolute emergency, the settings made for adjustment and calibration can be reset in full or in part to the factory settings and/or "Adjustment/calibration"...
  • Page 236: Adjusting The Touchscreen Via The Web Service

    Example: https://192.168.15.240 Tip: The IP address of your unit can be found on the info screen. 3. Confirm your input with the "Enter" key. The Sirona Web service is loaded. "User" The sub menus of the menu are displayed in the structure tree.
  • Page 237 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.3 Adjusting the touchscreen via the web service A message is displayed in the web browser to indicate that the adjustment screen has been opened on the display. The first adjustment screen appears on the control panel.
  • Page 238 7 Unit adjustment and calibration Sirona Dental Systems GmbH 7.3 Adjusting the touchscreen via the web service Service Manual ORTHOPHOS SL 8. Touch the center of the square in the bottom right corner of the screen. 9. Touch the center of the square in the bottom left corner of the screen.
  • Page 239 Sirona Dental Systems GmbH 7 Unit adjustment and calibration Service Manual ORTHOPHOS SL 7.3 Adjusting the touchscreen via the web service The unit switches to PC service mode. Service mode is displayed on the Easypad via the PC service image. 11. Exit the web service again.
  • Page 240: Service Routines

    Note on the web service service menu of the control panel. The service routines S008.3, S017.2 - S017.18 and S037.4 can also be run via the Sirona web service (see section Running service routines via the web service [ → 327]). 8.1.1...
  • Page 241 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel Service routine Function S009 Flash file system S. [ → 268] S009.4 Formatting flash file system S. [ → 268] S011 Dosimetry (without ring movement) S.
  • Page 242 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL Service routine Function S021 Service for the packing position S. [ → 304] S021.3 Initiate the packing position for the diaphragm S. [ → 302] S021.4...
  • Page 243: Alphabetical List Of Service Routine Functions

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel Alphabetical list of service routine functions 8.1.1.2 Function Service routine Inquiry of the CAN status register of the modules S012.2 S. [ → 273] Switch off cool-down interval of the single tank S005.6...
  • Page 244 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL Function Service routine Setting the default IP address, default gateway address and S037.2 S. [ → 322] default subnet mask Calibrating the occlusal bite block S018.9...
  • Page 245 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel Function Service routine Test of the ceph image path S033.10 S. [ → 307] Overview of module software versions S008.2* S. [ → 267]...
  • Page 246: Service Menu And Service Routines

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 8.1.2 Service menu and service routines You can use the service routines to check the function of certain unit components and modules, as well as to set important unit parameters.
  • Page 247 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel X RAY Radiation can be released. X RAY Active! Caution! Radiation is being released. Selection field 1 Display fields for service routines, test steps, values, unit parameters, etc.
  • Page 248: Basic Operating Procedures In The Service Menu Of The Control Panel

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 8.1.3 Basic operating procedures in the service menu of the control panel Calling the service menu 8.1.3.1 The service menu is intended exclusively for service engineers. Service routines can be activated and unit settings incl.
  • Page 249: Selecting Service Routines And Test Steps

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel Selecting service routines and test steps 8.1.3.2 8.1.3.2.1 Selecting a service routine ✔ The service menu must be selected. ➢ Select the desired service routine via the arrow keys in the selection field 1 (A) and confirm the selection via the service key (B).
  • Page 250 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 8.1.3.2.3 Service routines with security access A security code is required for accessing service routines involving functions such as radiation release or editing of configuration data or stored values.
  • Page 251: Select Parameters

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel Select parameters 8.1.3.3 If arrow keys are displayed in the selection fields once the required service routine has been selected, you can use these arrow keys to choose between different parameters.
  • Page 252: Exiting The Test Step And Service Routine

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL Exiting the test step and service routine 8.1.3.5 Touch the Service key (A) or the double arrow key (B) to go back to the menu for selecting service routines.
  • Page 253: S001: Radiation Without Rotary Movement, Fixed Maximum Radiation Time

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 8.1.4 S001: Radiation without rotary movement, fixed maximum radiation time SA** Function S001 X-ray beam test kV/ma levels test Explanation: ● Radiation with selected kV/mA level ●...
  • Page 254 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 3. Initiate the radiation. Radiation is released from any tube assembly position. There is no rotational movement. The primary diaphragm is in the last position set.
  • Page 255: S002: Radiation Without Rotary Movement, Selectable Kv/Ma Level 255 And Maximum Radiation Time 8.1.5.1 S002: Test Step 1

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 8.1.5 S002: Radiation without rotary movement, selectable kV/ mA level and maximum radiation time S002 SA** Function S002 X-ray beam test Explanation: ● Radiation with selected kV/mA level ●...
  • Page 256: S002: Test Step 3

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 3. Initiate the radiation. Radiation is released from any tube assembly position. There is no rotational movement. The primary diaphragm is in the last position set.
  • Page 257 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 2. Use the arrow keys (A) to select the required kV/mA level and the required radiation time (see table). 3. Initiate the radiation.
  • Page 258: S002: Test Step 4

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL S002: Test step 4 8.1.5.3 Radiation without rotation; maximum radiation time selectable; the primary diaphragm is set to a widened opening symmetrical to the PAN setting so that the step filter still remains covered.
  • Page 259 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel The radiation time corresponds to the selected radiation time. Important: If you let go of the release button before this radiation time expires, radiation will be terminated prematurely. The actual radiation time is not displayed.
  • Page 260: S005: General X-Ray Tube Assembly Service

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 8.1.6 S005: General X-ray tube assembly service SA** Function S005 General X-ray tube assembly service S005.2 Automatic adjustment of preheating S005.4 Fan test S005.5...
  • Page 261: S005: Test Step 4

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel During the adjustment procedure, a progress indicator is displayed in selection field 1. "CONTINUOUS" is displayed in selection field 2. After a successful adjustment, the offset value is displayed in selection field 3.
  • Page 262: S005: Test Step 5

    Running this service routine is irreversible! The device is operated beyond its specifications afterwards. This may result in an X-ray tube assembly failure. As a result, the system warranty expires. SIRONA is able to prove this system setting after a system failure. IMPORTANT Upon explicit customer request, the monitored pulse/pause ratio can be disabled in exceptional cases.
  • Page 263 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel If pulse/pause monitoring has already been disabled prematurely, this service routine simply displays the status and the arrow keys are not shown. The status cannot be changed.
  • Page 264: S005: Test Step 7

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL S005: Test step 7 8.1.6.5 Configuring continuous operation of the tube assembly fan NOTICE Contamination of the fan and the single tank housing...
  • Page 265: S007: Fault Buffer

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 8.1.7 S007: Fault buffer SA** Function S007 Fault buffer S007.1 View error memory S007.2 Clear error memory * SR=service routine, ** SA=security access 8.1.7.1...
  • Page 266: S007: Test Step 2

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL S007: Test step 2 8.1.7.2 Clear error memory 1. Call service routine S007.2 [ → 249]. The system's readiness to clear the memory is indicated by the display message "FFFF"...
  • Page 267: S008: Update Service

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 8.1.8 S008: Update service SA** Function S008 Update service S008.2 Overview of module software versions S008.3 Confirming the unit serial number * SR=service routine, ** SA=security access 8.1.8.1...
  • Page 268: S009: Flash File System

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 8.1.9 S009: Flash file system IMPORTANT When the flash file system is formatted, the content of the error logging memory is lost. IMPORTANT After running service routine SR 009.4 a software update to the currently...
  • Page 269: S011: Dosimetry (Without Ring Movement)

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 8.1.10 S011: Dosimetry (without ring movement) WARNING Unit is radiating X-rays. Excess exposure to X-rays is detrimental to health. ➢ Use the prescribed accessories for radiation protection.
  • Page 270 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL The kV/mA level is displayed in selection field 1. The corresponding radiation time is displayed in selection field 2. 3. Use the arrow keys (A) to select the required kV/mA level, e.g. "85kV/ 5mA".
  • Page 271: S011: Test Step 24

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S011: Test step 24 8.1.10.2 Dosimetry with pulsed radiation A dosimeter for pulsed radiation (e.g. Mult-O-Meter 512L) is required for dosimetry. 1. Attach the Mult-O-Meter sensor in the middle of the sensor unit (flat panel detector side).
  • Page 272: S012: Can Bus Service

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 8.1.11 S012: CAN bus service SA** Function S012 CAN bus service S012.1 Presence display of modules S012.2 Inquiry of the CAN status register of the modules S012.3...
  • Page 273: S012: Test Step 2

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S012: Test step 2 8.1.11.2 Inquiry of the CAN status register of the modules IMPORTANT Before querying the CAN status register for the modules, you should first run service routine S012.3 to reset the registers [ →...
  • Page 274: S012: Test Step 3

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL S012: Test step 3 8.1.11.3 Resetting the CAN status register of the modules IMPORTANT Run this service routine as required before service routine S012.2.
  • Page 275: S014: Rotation Motor Service

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 8.1.12 S014: Rotation motor service SA** Function S014 Rotation motor service S014.3 Free travel of rotational drive S014.4 Display of light barrier signals of rotational drive * SR=service routine, ** SA=security access 8.1.12.1...
  • Page 276: S015: Actuator Service

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 8.1.13 S015: Actuator service SA** Function S015 Actuator service S015.5 Functional test of actuators 1 + 2 * SR=service routine, ** SA=security access 8.1.13.1...
  • Page 277: S017: Configuration Service

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 8.1.14 S017: Configuration service NOTE: The unit can also be configured via the web service via service routine [ → 327] S017. IMPORTANT After making changes to the unit configuration, error message E71130 appears on the control panel's touchscreen.
  • Page 278: S017: Test Step 2

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL S017: Test step 2 8.1.14.1 Configuring the hardware version Selection Unit variants Function field PanDCS__3D_______ Panoramic digital, 2D/3D-Sensor PanDCS__3D_CephL Panoramic digital, 2D/3D-Sensor, Ceph left digital...
  • Page 279 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel Checking the jumper position on board DX1 ➢ Check the jumper positions of the sockets on board DX1: X309 (DX91, ceph sensor) and...
  • Page 280 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL Without ceph Jumper inside: Cephalometer not connected, i.e. connector not plugged Ceph left or Ceph right Jumper outside: Cephalometer connected, i.e. connector plugged in.
  • Page 281: S017: Test Step 3

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S017: Test step 3 8.1.14.2 Enter the country group code Selection field Code Function Worldwide* Asia * Factory setting 1. Call service routine S017.3.
  • Page 282: S017: Test Step 4

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL S017: Test step 4 8.1.14.3 Select a language Selection field Code Function* English* German French Italian Dutch Spanish Russian Portuguese Chinese (PRC) Korean...
  • Page 283: S017: Test Step 6

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S017: Test step 6 8.1.14.4 Activate / deactivate the remote control display Selection field Code Function Display inactive* Display active * factory settings 1.
  • Page 284: S017: Test Step 8

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL S017: Test step 8 8.1.14.5 Selecting the kV/mA level series Selection field Code Function 16 mA series for pan 16 mA series for Ceph...
  • Page 285: S017: Test Step 13

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S017: Test step 13 8.1.14.6 Enable/disable the welcome screen Selection field Code Function Welcome screen disabled Welcome screen enabled* * Factory setting 1. Call service routine S017.13.
  • Page 286: S017: Test Step 14

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL S017: Test step 14 8.1.14.7 Enable/disable certain lines of the welcome screen Selection field Code Meaning/Function First name Surname Date of birth Patient number...
  • Page 287: S017: Test Step 15

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S017: Test step 15 8.1.14.8 Selection field Code Function Acoustic signal indicating the end of the exposure is disabled Acoustic signal indicating the end of the...
  • Page 288: S017: Test Step 18

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL S017: Test step 18 8.1.14.9 Occlusal bite block function The occlusal bite block function can be activated or deactivated via service routine S017.18.
  • Page 289: S018: Service For Height Adjustment

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 8.1.15 S018: Service for height adjustment SA** Function S018 Service for height adjustment S018.2 Set the maximum travel height S018.3 Undo the maximum travel height setting S018.4...
  • Page 290: S018: Test Step 3

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 4. *For greater clarity the unit is shown in the adjacent image without a cover. Set the mechanical limit stop on the unit: Loosen the nut (D) and move mechanical limit stop (E) for the limit switch until it engages.
  • Page 291: S018: Test Step 4

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S018: Test step 4 8.1.15.3 Check the height adjustment sensor system This service routine is used to move the unit up or down as far as the limit switches using the Up/Down keys on the control panel.
  • Page 292: S018: Test Step 5

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL S018: Test step 5 8.1.15.4 Setting the minimum travel height Setting the minimum travel height is possible only for a unit height that is below the lower correction switch level (<...
  • Page 293: S018: Test Step 7

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S018: Test step 7 8.1.15.6 NOTE: In order to execute this service routine, the occlusal bite block First execute SR S017.18 function must be activated via service routine S017.18[ → 288].
  • Page 294 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 2. Use the arrow keys (H) in selection field 1 to select the required program group. 3. Use the arrow keys (I) to select the required nominal angle (possible range: -17°...
  • Page 295: S018: Test Step 8

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S018: Test step 8 8.1.15.7 NOTE: In order to execute this service routine, the occlusal bite block First execute SR S017.18 function must be activated via service routine S017.18[ → 288].
  • Page 296 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 8.1.15.7.1 Codes of possible error modes Display Error Troubleshooting measures Normal state No action Plug X1000 is not plugged in on Check plug for occlusal bite block on board DX1.
  • Page 297: S018: Test Step 9

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S018: Test step 9 8.1.15.8 NOTE: In order to execute this service routine, the occlusal bite block First execute SR S017.18 function must be activated via service routine S017.18[ → 288].
  • Page 298 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL U: Adjusting and saving the upper value 1. Insert the occlusal bite block into the bite block holder and deflect the bite block plate all the way up (up to the stop).
  • Page 299: S018: Test Step 10

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S018: Test step 10 8.1.15.9 NOTE: In order to execute this service routine, the occlusal bite block First execute SR S017.18 function must be activated via service routine S017.18[ → 288].
  • Page 300: S020: Service For Temple Support

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 8.1.16 S020: Service for temple support SA** Function S020 Service for temple support S020.1 Programming the switching thresholds of the temple support * SR=service routine, ** SA=security access 8.1.16.1...
  • Page 301 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 2. Use the arrow keys (A) in the corresponding selection field to select the required offset for the switching point (see table). IMPORTANT A negative offset shifts the respective switching point toward larger skulls;...
  • Page 302: S021: Service For The Packing Position

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 8.1.17 S021: Service for the packing position SA** Function S021 Service for packing position S021.3 Initiate the packing position for the diaphragm S021.4...
  • Page 303: S021: Test Step 4

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S021: Test step 4 8.1.17.2 Initiate the packing position for the sensor unit 1. Call service routine S021.4 [ → 249]. An inactive progress indicator in selection field 1 and the message FFFF in selection field 2 signal that the system is ready to move the sensor unit.
  • Page 304: S032: Sensor Test

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 8.1.18 S032: Sensor test SA** Function S032 Sensor test S032.10 Self-test of the DCS image data path S032.50 Self-test of the FP image data path 8.1.18.1...
  • Page 305: S032: Test Step 50

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S032: Test step 50 8.1.18.2 Self-test of the FP image data path 1. Call service routine S032.50 [ → 249]. Once the service routine has been selected, "FP test" is displayed in selection field 1.
  • Page 306: Possible Results Of Self-Test And Troubleshooting Measures

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL Possible results of self-test and troubleshooting measures 8.1.18.4 "Success" The self-test was successfully executed. No errors were detected in the image data path.
  • Page 307: S033: Test Of Ceph Image Path (For Ceph Units Only)

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 8.1.19 S033: Test of ceph image path (for ceph units only) SA** Function S033 Test of the ceph image path S033.10 Test of the ceph image path * SR=service routine, ** SA=security access 8.1.19.1...
  • Page 308 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL The result of the synchronized readout sequence is displayed (see S032: Test step 10: Result of synchronized readout sequence. 4. Exit the service routine [ → 252].
  • Page 309: S034: Service For The Digital Cephalometer

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 8.1.20 S034: Service for the digital cephalometer SA** Function S034 Service for the digital cephalometer S034.4 Display/calibrate center position of Ceph scan sensor axis S034.5...
  • Page 310 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL In the test mode, travel to the center position of the ceph scan sensor axis is initially performed without offsets. The T key lights up and indicates the test mode.
  • Page 311 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 312 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL Room side Wall side Ceph arm mounted on left-hand side Ceph arm mounted on right-hand side Sensor side 8. Calculate the offset value according to the following formula: For left-side arm: (L2 - L1) ÷...
  • Page 313: S034: Test Step 5

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S034: Test step 5 8.1.20.2 Center position of ceph scan secondary diaphragm axis, display/ calibrate 1. If not already done, detach the cover to the cephalometer [ → 37].
  • Page 314 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 315 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel Room side Wall side Ceph arm mounted on left-hand side Ceph arm mounted on right-hand side Secondary diaphragm side 7. Calculate the offset value according to the following formula: For left-side arm: (L2 - L1) ÷...
  • Page 316: S034: Test Step 6

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL S034: Test step 6 8.1.20.3 Moving to the Ceph packing position 1. Call service routine S034.6 [ → 249]. After calling the service routine, an inactive progress indicator is displayed in selection field 1.
  • Page 317: S034: Test Step 8

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S034: Test step 8 8.1.20.4 User-specific determination of the shadowing limit for C1 and C2 For ceph programs C1 and C2, the default setting of the shadowing limit can be assigned an individual offset.
  • Page 318: S034: Test Step 9

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 2. Select the desired offset value using the arrow keys (A) in selection field 1 (step width = 500, offset value = 1000 corresponds to a 1 mm shift of the shadowing limit).
  • Page 319 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel The offset must be determined by reference to an existing exposure. Based on the default setting, an offset in the range of +15,000 μm to - 30,000 μm can be selected and saved.
  • Page 320: S037: Network Service

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 8.1.21 S037: Network service SA** Function S037 Network service S037.1 Displaying the network data S037.2 Setting the default IP address, default gateway address and default subnet mask S037.3...
  • Page 321 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 1. Call service routine S037.1. Once the service routine has been selected, the IP address of the unit is displayed in selection field 1 (see table).
  • Page 322: S037: Test Step 2

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL S037: Test step 2 8.1.21.2 Setting the default IP address, default gateway address and default subnet mask IMPORTANT The network address can only be restored to the factory setting (default value) in fixed address boot mode (STATIC or no DHCP).
  • Page 323: S037: Test Step 3

    Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel S037: Test step 3 8.1.21.3 Configuring boot mode Selection field Parameters Meaning DYNAMIC Automatic address assignment (DHCP/AutoIP) STATIC Fixed address* * Factory setting 1.
  • Page 324: S037: Test Step 4

    8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL S037: Test step 4 8.1.21.4 Manual input of static network settings (IP address, default gateway address, and subnet mask) This service routine cannot run in DYNAMIC mode (T key is blocked).
  • Page 325 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.1 Perform service routines via the control panel 1. Call service routine S037.4. Once the service routine has been selected, the IP address of the unit is displayed in selection field 1.
  • Page 326 8 Service routines Sirona Dental Systems GmbH 8.1 Perform service routines via the control panel Service Manual ORTHOPHOS SL 5. Use the arrow keys (F) to select the digit to be changed in selection field 1 ("Digit No. 12" in the example).
  • Page 327: Running Service Routines Via The Web Service

    8.2 Running service routines via the web service Running service routines via the web service NOTE: The web service is only available in English NOTE: The following service routines can be run using the Sirona Web service: "Service" In the main menu ●...
  • Page 328 Sirona Dental Systems GmbH 8.2 Running service routines via the web service Service Manual ORTHOPHOS SL "Continue to this Website" 4. Click on the link The Sirona Web service is loaded. "User" The sub menus of the menu are displayed in the structure tree.
  • Page 329 Sirona Dental Systems GmbH 8 Service routines Service Manual ORTHOPHOS SL 8.2 Running service routines via the web service "Service" The sub menus of the menu are displayed in the structure tree. 8. Select the desired service routine in the structure tree, e.g. S008.3.
  • Page 330: Repair

    For reasons of product safety, this product should be operated only with original accessories from Sirona or Sirona-approved accessories from third parties. The user is responsible for any damage resulting from the use of non-approved accessories.
  • Page 331: Safety Checks

    Measures following replacement of boards. Also check whether the current software CD or the SIRONA dealer page contains any additional more up-to-date information about module replacement. Be sure to follow the instructions about how to proceed following module replacement.
  • Page 332: Height Adjustment Motor (M1_4)/Spindle

    You must lock the slide in position to prevent the risk of the carriage plummeting downward during service work if the self-locking function does not fully engage. Sirona recommends using the height adjustment service kit, REF. 62 57 518. This service kit is used to prevent automatic downward movement of the slide during service by fixing the slide and the spindle holder.
  • Page 333 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.2 Height adjustment motor (M1_4)/spindle 9.2.1.1.1 Moving the slide with the "height adjustment" service kit, REF. 62 57 518 move the slide 1. Insert the clamp through the opening (A) in the stand and rotate it 90°.
  • Page 334 9 Repair Sirona Dental Systems GmbH 9.2 Height adjustment motor (M1_4)/spindle Service Manual ORTHOPHOS SL 9.2.1.1.2 Moving the slide without the "height adjustment" service kit, REF. 62 57 518 move the slide 1. Loosen the two screws (C) and remove the cover (D).
  • Page 335: Removing Board Dx32

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.2 Height adjustment motor (M1_4)/spindle 9.2.2 Removing board DX32 1. Unscrew the bracket (C). The bracket (C) is located in front of board DX32. 2. Move the stand to a height of 1260 (control panel display) 3.
  • Page 336 9 Repair Sirona Dental Systems GmbH 9.2 Height adjustment motor (M1_4)/spindle Service Manual ORTHOPHOS SL 8. The cable L3 can stay on the top covering plate (E). 9. Remove connector X2 from board DX32 and remove the protective conductor. 10. Remove cable L2 from terminal X100 and pull it downwards from out of the connection box (F).
  • Page 337: Replacing The Height Adjustment Motor/Spindle

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.2 Height adjustment motor (M1_4)/spindle 9.2.3 Replacing the height adjustment motor/spindle Removing the spindle 1. Loosen the two screws (E) on the spindle holder (F) (if you have not already done so) [ → 332].
  • Page 338 9 Repair Sirona Dental Systems GmbH 9.2 Height adjustment motor (M1_4)/spindle Service Manual ORTHOPHOS SL Installing the new motor Install the height adjustment motor in the reverse order of removal. Please observe the following: Nuts: When fastening the motor, make sure that all three screws are tightened uniformly and protrude approx.
  • Page 339: Laying Of Cables When Replacing The Height Adjustment Motor

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.2 Height adjustment motor (M1_4)/spindle 9.2.4 Laying of cables when replacing the height adjustment motor 1. Plug connector X2 (O) into board DX32. 2. Connect the protective ground wire (P) and lay it as shown in the photo.
  • Page 340 9 Repair Sirona Dental Systems GmbH 9.2 Height adjustment motor (M1_4)/spindle Service Manual ORTHOPHOS SL 3. Attach cable L2 first to the lower strain relief (photo on the left) and then to the upper strain relief (photo on the right) (Q) of board DX32.
  • Page 341: What Has To Be Done After Replacing The Height Adjustment Motor

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.2 Height adjustment motor (M1_4)/spindle 6. Lay the motor cable in the cable harness (V) on the rear of the unit and secure in position with the cable clamps. 7. Route the cable into the arm.
  • Page 342: Ring Motor (M1_3)

    9 Repair Sirona Dental Systems GmbH 9.3 Ring motor (M1_3) Service Manual ORTHOPHOS SL Ring motor (M1_3) 9.3.1 Replacing the ring motor Removing the covers ➢ Remove the "arm cover" [ → 37]. Removing the defective motor 1. Detach the motor cable from the cable harness and pull it off of connector X813 on board DX1.
  • Page 343 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.3 Ring motor (M1_3) Replacing the pinion 1. Loosen the set screws (A) and pull off the defective pinion (B). 2. IMPORTANT: Ensure that the pinion is seated in the coupling so that the set screws (A) are sitting on the flattened surface (C) of the pinion during subsequent tightening to prevent the pinion from turning.
  • Page 344: Laying Of Cables When Replacing The Ring Motor

    9 Repair Sirona Dental Systems GmbH 9.3 Ring motor (M1_3) Service Manual ORTHOPHOS SL 9.3.3 Laying of cables when replacing the ring motor 1. Lay the cable (G) in a loop from top to bottom and fix it in the cable clamp (J).
  • Page 345: Pan Actuators (M1_1/2)

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.4 Pan actuators (M1_1/2) Pan actuators (M1_1/2) 9.4.1 Replacing actuators Removing the covers ➢ Remove the "arm cover, top" [ → 37]. Removing the defective actuator 1. Detach the actuator cable from the cable holders and pull it off connector X812 (AK2;...
  • Page 346: Laying Of Cables When Replacing The Actuator

    9 Repair Sirona Dental Systems GmbH 9.4 Pan actuators (M1_1/2) Service Manual ORTHOPHOS SL 9.4.2 Laying of cables when replacing the actuator ➢ Lay the cables in a loop and secure the loop with the cable ties (A). 9.4.3 What has to be done after replacing the actuators? 1.
  • Page 347: Head Support

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.5 Head support Head support 9.5.1 Replacing the headrest Removing the defective headrest 1. Hold the headrest firmly from below, loosen screw (A), and remove the defective headrest. Tip: To remove the headrest, this must be tilted slightly.
  • Page 348: Laying Of Cables When Replacing The Headrest

    9 Repair Sirona Dental Systems GmbH 9.5 Head support Service Manual ORTHOPHOS SL 9.5.2 Laying of cables when replacing the headrest ➢ Plug cable L13 into connector X1 on board DX5. 9.5.3 What has to be done after replacing the headrest? ➢ Use the forehead and temple support keys on the touchscreen to check the function of the headrest.
  • Page 349: Control Panel

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.6 Control panel Control panel 9.6.1 Replacing the user interface 1. Press into slit (A) of the housing cover with a screwdriver (do not pry!) and remove the defective user interface from the control panel.
  • Page 350: What Has To Be Done After Replacing The User Interface

    9 Repair Sirona Dental Systems GmbH 9.6 Control panel Service Manual ORTHOPHOS SL 5. Update the nameplate at the control panel cover. To do so, affix the supplied label as shown in the figure. 9.6.1.1 What has to be done after replacing the user interface?
  • Page 351: Replacing The Control Panel

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.6 Control panel 9.6.2 Replacing the control panel 1. Remove the user interface [ → 349]. 2. Swing the folding mechanism for the Easypad (A) all the way up. This makes it easier to remove the cover.
  • Page 352: What Has To Be Done After Replacing The Control Panel

    9 Repair Sirona Dental Systems GmbH 9.6 Control panel Service Manual ORTHOPHOS SL What has to be done after replacing the control panel? 9.6.2.1 IMPORTANT "ORTHOPHOS Use the most current software version available on the SL Firmware DVD" 1. Switch the unit on.
  • Page 353: Laser Light Localizers

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.7 Laser light localizers Laser light localizers CAUTION Risk of injury to eyes. The unit contains lasers of Class 1. Keep a distance of at least 4" (10 cm) between eye and laser. Do not look into the laser beam.
  • Page 354: Adjusting The Laser Light Localizer Fh (Pan)

    9 Repair Sirona Dental Systems GmbH 9.7 Laser light localizers Service Manual ORTHOPHOS SL 6. Loosen the screws (B) and carefully pull the laser module (C) out of the holder toward the front. 7. Insert the new laser module (FH) in the holder and retighten screws (B).
  • Page 355 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.7 Laser light localizers 6. Rotate the mirror by pressing into the left depression (D) of the toolbar. 7. Touch the light localizer key (E) on the touchscreen. Then light localizers are switched on.
  • Page 356: Fh Laser Light Localizer (Ceph)

    9 Repair Sirona Dental Systems GmbH 9.7 Laser light localizers Service Manual ORTHOPHOS SL 9.7.2 FH laser light localizer (Ceph) Replacing the ceph laser module in the FH laser light localizer 9.7.2.1 (ceph) 1. Remove the covers "Cephalometer cover" and "Outside secondary diaphragm"...
  • Page 357: Adjusting The Fh Laser Light Localizer (Ceph)

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.7 Laser light localizers Adjusting the FH laser light localizer (Ceph) 9.7.2.2 ✔ Reattach the "Outside secondary diaphragm" cover [ → 37]. 1. Switch the unit on via the switch (A) (see also Operating Instructions).
  • Page 358 9 Repair Sirona Dental Systems GmbH 9.7 Laser light localizers Service Manual ORTHOPHOS SL 6. Align the Ceph (FH) laser module so that FH light beam (D) runs horizontally between the ear plugs (tolerance: ± 1.5 mm). 7. NOTICE! Ensure that you do not turn the laser module while tightening the screws.
  • Page 359: Ms Laser Light Localizer (Pan)

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.7 Laser light localizers 9.7.3 MS laser light localizer (pan) Replacing the laser module in the MS laser light localizer (pan) 9.7.3.1 1. Remove the "Top arm" cover [ → 37]. 2. Pull the cable of the MS laser module off connector X811 on board DX1.
  • Page 360: Adjusting The Ms Laser Light Localizer (Pan)

    9 Repair Sirona Dental Systems GmbH 9.7 Laser light localizers Service Manual ORTHOPHOS SL Adjusting the MS laser light localizer (PAN) 9.7.3.2 ✔ The "Front slide" cover must have been removed [ → 37]. 1. Insert the forehead and temple supports (see also Operating Instructions).
  • Page 361 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.7 Laser light localizers 6. Slightly loosen the screws (F) of the laser module and align the laser module (MS) so that the MS light beam is displayed at the center of the bite block or bite block holder.
  • Page 362: Laser Area Display (Vol)

    9 Repair Sirona Dental Systems GmbH 9.7 Laser light localizers Service Manual ORTHOPHOS SL 9.7.4 Laser area display (VOL) Replacing the laser area display (VOL) 9.7.4.1 1. Remove the "Top arm" and "Front slide" covers [ → 37]. 2. Remove the cable of the defective laser area display (C) from connector X1500 on board DX1 and loosen the cable from the cable holders.
  • Page 363 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.7 Laser light localizers 5. Ensure that the threads (B) on the new laser area display are unscrewed so far that the notch is released (A). 6. NOTICE! Ensure that the laser area display is attached to the cast part and that there is no gap between the laser area display and the cast part.
  • Page 364: Adjusting The Laser Area Display (Vol)

    9 Repair Sirona Dental Systems GmbH 9.7 Laser light localizers Service Manual ORTHOPHOS SL Adjusting the laser area display (VOL) 9.7.4.2 1. Switch the unit on via the switch (A) (see also Operating Instructions). The X-ray radiation indicator (B) lights up briefly.
  • Page 365 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.7 Laser light localizers If threads B should tightened: Slightly loosen the screws (B) of the laser area display. 7. Adjust the laser area display (A) vertically in such a way that the distance between the displayed laser line and cover joint is 5 mm.
  • Page 366: Checking The Connection Of The Ms And Fh (Pan) Laser Light

    9 Repair Sirona Dental Systems GmbH 9.7 Laser light localizers Service Manual ORTHOPHOS SL 9.7.5 Checking the connection of the MS and FH (pan) laser light localizers NOTE: The cables/plugs of the laser light localizers are not labeled ex works. Should the sockets for plugs X607 (MS) and X611 (FH) get mixed up on board DX1, there will be a control error in the 3D design of the ORTHOPHOS SL X-ray unit.
  • Page 367: Occlusal Bite Block

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.8 Occlusal bite block Occlusal bite block 9.8.1 Replacing the occlusal bite block ➢ Replace the defective occlusal bite block with the new one. 9.8.2 What has to be done after replacing the occlusal bite block? 1.
  • Page 368: Bite Block Holder

    9 Repair Sirona Dental Systems GmbH 9.9 Bite block holder Service Manual ORTHOPHOS SL Bite block holder 9.9.1 Replacing the bite block holder 1. Loosen the two screws (A) and remove the bite block holder. 2. Unplug the connector of cable L41 from board DX51.
  • Page 369: Replacing The Silicone Rings

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.9 Bite block holder 9.9.2 Replacing the silicone rings See also the chapter "Replacing the bite block holder [ → 368]". 1. Loosen the two screws (A) and remove the bite block holder.
  • Page 370: Diaphragm Unit

    9 Repair Sirona Dental Systems GmbH 9.10 Diaphragm unit Service Manual ORTHOPHOS SL 9.10 Diaphragm unit 9.10.1 Replacing the diaphragm unit IMPORTANT Only the diaphragm unit of ORTHOPHOS SL may be installed! Note the dependency of the diaphragm unit on device serial numbers in the spare parts list.
  • Page 371: What Has To Be Done After Replacing The Diaphragm Unit

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.10 Diaphragm unit 9.10.2 What has to be done after replacing the diaphragm unit? IMPORTANT "ORTHOPHOS Use the most current software version available on the SL Firmware DVD" 1. Switch the unit on.
  • Page 372: X-Ray Tube Unit

    9 Repair Sirona Dental Systems GmbH 9.11 X-ray tube unit Service Manual ORTHOPHOS SL 9.11 X-ray tube unit DANGER Perilous shock hazard! It is essential to switch the unit off and to wait at least another 4 minutes before starting the repair or taking off a cover panel! 9.11.1...
  • Page 373 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.11 X-ray tube unit 5. Loosen the two rear screws (C) on the tube assembly. 6. CAUTION! The tube assembly is heavy! Hold the tube assembly firmly in place, loosen the two front screws D (3-4 turns) and remove the tube assembly toward the front.
  • Page 374: Replacing The Fan In The X-Ray Tube Assembly

    9 Repair Sirona Dental Systems GmbH 9.11 X-ray tube unit Service Manual ORTHOPHOS SL 9.11.2 Replacing the fan in the X-ray tube assembly Replacing the fan 9.11.2.1 IMPORTANT As the X-ray tube assembly has to be disassembled when replacing the fan, the unit must be fully readjusted (recalibrated) [ → 160] after replacing the fan.
  • Page 375: Cables And Connectors For Replacement Of The X-Ray Tube Assembly

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.11 X-ray tube unit 9.11.3 Cables and connectors for replacement of the X-ray tube assembly 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 376 9 Repair Sirona Dental Systems GmbH 9.11 X-ray tube unit Service Manual ORTHOPHOS SL Cable L5 → socket J6 on board DX6 Cable L6 → socket J2/J3 on board DX6 Ground cable → Connector X304 (tube assembly 1.0) or X103 (tube assembly 2.0) on board DX6 Cable L3 →...
  • Page 377: What Has To Be Done After Replacing The X-Ray Tube Assembly

    IMPORTANT: The unit serial number is located on the nameplate of the unit. NOTICE! If the serial number is incorrect, cancel the update and contact the Sirona Customer Service Center! 6. Confirm the serial number using service routine S008.3. 7. Perform a software update to the current software version.
  • Page 378: Sensor Unit

    9 Repair Sirona Dental Systems GmbH 9.12 Sensor unit Service Manual ORTHOPHOS SL 9.12 Sensor unit NOTICE Damage to the Flat Panel Detector The Flat Panel Detector is a sensitive component. Be careful not to touch the detector surface during assembly. NOTICE Damage to the rotation unit The sensor is rotated via a motor drive.
  • Page 379 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.12 Sensor unit 4. Loosen the 3 screws (G) and unhinge the flat panel detector from the guide of the rotation unit. 5. Install the new flat panel detector in the reverse order of removal.
  • Page 380: Connecting A Sensor Unit

    9 Repair Sirona Dental Systems GmbH 9.12 Sensor unit Service Manual ORTHOPHOS SL Connecting a sensor unit 9.12.1.2 1. Insert lead L84 from the flat panel detector, as shown in the illustration, into plugs X601 and X602 on board DX83. 2. Lay the leads as shown in the illustration.
  • Page 381: What Has To Be Done After Replacing The Flat Panel

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.12 Sensor unit What has to be done after replacing the flat panel detector? 9.12.1.3 IMPORTANT "ORTHOPHOS Use the most current software version available on the SL Firmware DVD" 1. Reattach the covers of the sensor unit in reverse order of removal.
  • Page 382: Dcs Sensor

    9 Repair Sirona Dental Systems GmbH 9.12 Sensor unit Service Manual ORTHOPHOS SL 9.12.2 DCS sensor Replacing the DCS sensor 9.12.2.1 ✔ There is a backup file of the sensor data. NOTE: If there is no backup file for the sensor data, the sensor data must be downloaded from board DX83 and saved before replacing the DCS sensor (see the chapter titled "Backing up sensor data...
  • Page 383 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.12 Sensor unit 4. Loosen the 4 screws (B) and unhinge the DCS sensor from the lugs (A) of the rotation unit. 5. Install the new sensors in the reverse order of removal. For the instructions on connecting the sensor unit, see chapter "Connecting...
  • Page 384: Connecting A Sensor Unit

    9 Repair Sirona Dental Systems GmbH 9.12 Sensor unit Service Manual ORTHOPHOS SL Connecting a sensor unit 9.12.2.2 1. Insert leads L83A and L83B from the rotation unit, as shown in the illustration, into plugs X100, X700 and X800 on board DX83.
  • Page 385: What Has To Be Done After Replacing The Dcs Sensor

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.12 Sensor unit What has to be done after replacing the DCS sensor? 9.12.2.3 IMPORTANT "ORTHOPHOS Use the most current software version available on the SL Firmware DVD" 1. Reattach the covers of the sensor unit in reverse order of removal.
  • Page 386: Replace Board Dx83

    9 Repair Sirona Dental Systems GmbH 9.12 Sensor unit Service Manual ORTHOPHOS SL 9.12.3 Replace board DX83 ✔ There is a backup file of the sensor data. NOTE: If there is no backup file for the sensor data, the sensor data must be downloaded and saved from the board before replacing the DX83 board (see section “Backing up sensor data [ →...
  • Page 387 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.12 Sensor unit 4. Loosen the 4 screws (B) and unhinge the DCS sensor from the lugs (A) of the rotation unit. 5. Pull the fan package line off connector X1000 from board DX83 .
  • Page 388 9 Repair Sirona Dental Systems GmbH 9.12 Sensor unit Service Manual ORTHOPHOS SL 9. Hold the DCS sensor in place on the frame (K) and carefully remove the heatsink (L). NOTE: This can stick slightly to the DX83 board due to the two thermal conducting foils.
  • Page 389 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.12 Sensor unit 14. Place the two new heat conducting foils (Q) (included in the scope of supply for board DX83) on board DX83. 15. Reattach the heatsink with the four screws (J) and plain washers on the DCS sensor.
  • Page 390: Replacing Fan Package

    9 Repair Sirona Dental Systems GmbH 9.12 Sensor unit Service Manual ORTHOPHOS SL 9.12.4 Replacing fan package 1. Remove the cover of the sensor unit. 2. Loosen the cable of the defective fan package (B) from the cable holders on the sensor unit and pull it out from connector X1000 on board DX83.
  • Page 391: Cephalometer

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.13 Cephalometer 9.13 Cephalometer 9.13.1 Replacing the ceph sensor holder 1. Remove the sensor (A). 2. NOTICE! Do not tear off cable L35. Loosen the 3 screws (B) and remove the connection socket (C).
  • Page 392: Replacing The Ceph Sensor

    9 Repair Sirona Dental Systems GmbH 9.13 Cephalometer Service Manual ORTHOPHOS SL 9.13.2 Replacing the ceph sensor Removal and insertion of the ceph sensor is described in the Operating Instructions of the unit. 9.13.2.1 What has to be done after replacing the ceph sensor? IMPORTANT "ORTHOPHOS...
  • Page 393: Light Barriers

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.14 Light barriers 9.14 Light barriers 9.14.1 Replacing the light barriers The following light barriers can be replaced (see also Spare parts list): X-ray unit V1_1 X802 X803 V1_2 X811 X812 ● Light barrier at actuator 1, ON position: V1_1 ●...
  • Page 394: What Has To Be Done After Replacing The Light Barriers

    9 Repair Sirona Dental Systems GmbH 9.14 Light barriers Service Manual ORTHOPHOS SL Cephalometer V91_2 V91_1 ● Light barrier for patient diaphragm position detection, ceph: V91_1 ● Light barrier for sensor position detection, ceph: V91_2 9.14.2 What has to be done after replacing the light barriers? After replacing V1_1, V1_2 or V1_3 ➢...
  • Page 395: Boards

    "ORTHOPHOS SL Firmware DVD" and on the Sirona dealer page on the Internet. For this reason, you should always check for the latest information on the replacement of modules and performing updates before you start replacing any modules or boards.
  • Page 396: Replacing Boards

    9 Repair Sirona Dental Systems GmbH 9.15 Boards Service Manual ORTHOPHOS SL 9.15.2 Replacing boards CAUTION Risk of damage to boards Please observe the usual precautionary measures for handling printed circuit boards (ESD). Touch a ground point to discharge static electricity before touching any boards.
  • Page 397: Replacing Board Dx11

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.15 Boards Replacing board DX11 9.15.2.2 ✔ The cover plate of board DX11 must be removed [ → 396]. 1. CAUTION! Touch a ground point to discharge static electricity before touching any boards.
  • Page 398: Measures Following Replacement Of Boards

    "ORTHOPHOS SL Firmware The current update file is located on the DVD" on the Sirona dealer page on the Internet. 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 399 IMPORTANT: The unit serial number is located on the nameplate of the unit. NOTICE! If the serial number is incorrect, cancel the update and contact the Sirona Customer Service Center! 11. Confirm the serial number using service routine S008.3 [ → 267]. 12. Switch the unit off.
  • Page 400: Cable

    9 Repair Sirona Dental Systems GmbH 9.16 Cable Service Manual ORTHOPHOS SL 9.16 Cable 9.16.1 Replacing energy chain 1 completely Removing the defective energy chain IMPORTANT Remove energy chain Remove cable ties For the following steps, all necessary cable ties should be removed with wire cutters.
  • Page 401 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.16 Cable 4. If not present: Mark the position of the screw (B) on the stand. 5. Make a note of the position of the screw (B) for when you later install the new energy chain.
  • Page 402 9 Repair Sirona Dental Systems GmbH 9.16 Cable Service Manual ORTHOPHOS SL 9. Remove the cable covers (I) and (J) of the right-hand cable duct. 10. Remove the energy chain along with the mains filter plate from the stand. Installing the new energy chain 1.
  • Page 403 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.16 Cable IMPORTANT Possible assembly errors The one end piece has been removed to make the new energy chain easier to lay . The missing end piece is enclosed. ➢ Attach this end piece to the energy chain only when the energy chain has been laid in the stand.
  • Page 404: Replacing Cables

    9 Repair Sirona Dental Systems GmbH 9.16 Cable Service Manual ORTHOPHOS SL 9.16.2 Replacing cables CAUTION Switch the unit off before you start replacing cables or removing connectors. NOTICE Be careful not to twist the cables or kink the fiber-optic light guides when installing them.
  • Page 405: Replacing Fiber-Optic Cable L5, L6 Or L15

    Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.16 Cable Replacing fiber-optic cable L5, L6 or L15 9.16.2.1 1. Remove the defective fiber-optic cable. 2. NOTICE! Do not kink or twist fiber-optic cables, the bending radius may not be less than 20mm, otherwise it is at risk of breaking! Clip the two parts of the first radius limiter (A) close to the connector, which is plugged onto board DX1 onto the cable.
  • Page 406: Cable Exchange (L3, L5, L6, L11, L12, And L15)/Laying

    9 Repair Sirona Dental Systems GmbH 9.16 Cable Service Manual ORTHOPHOS SL Cable exchange (L3, L5, L6, L11, L12, and L15)/Laying the cable/ 9.16.2.2 corrugated tube at the rotation unit NOTICE The connectors and cables must be protected by inserting them in the fabric tube (A) supplied with the cables.
  • Page 407 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.16 Cable 9. Slide the corrugated tube over the spiral spring. 10. Remove the fabric tube and the pull wire. Laying the corrugated tube or cable at 1. Lay the corrugated tubes and cables back in their original position.
  • Page 408: Replacing Cable L117 Or L108 In Cable Track 2

    9 Repair Sirona Dental Systems GmbH 9.16 Cable Service Manual ORTHOPHOS SL Replacing cable L117 or L108 in cable track 2 9.16.2.3 1. Switch the unit on. 2. Move the slide downward to a pleasant working position using the Up/ Down keys on the control panel.
  • Page 409 Sirona Dental Systems GmbH 9 Repair Service Manual ORTHOPHOS SL 9.16 Cable 14. Remove the motor-side end piece from the cable track. 15. With defective cable L117: Unscrew cable L117 from the interface board and remove the shield. If cable L117 should be intact and used again, this step is not necessary.
  • Page 410: Replacing Cable L1 Or Grounding Strap In Cable Track 1

    9 Repair Sirona Dental Systems GmbH 9.16 Cable Service Manual ORTHOPHOS SL 21. CAUTION! Push the green cable. The white cable is carried along. In this way, you can prevent the sensitive fiber optic cable from being damaged. Push both cables (together) into the cable track up to the cable markings.
  • Page 411: Maintenance

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.1 Calibrating the unit Maintenance DANGER Potentially lethal shock hazard! It is essential to switch the unit off and to wait at least 1 minute, or 4 minutes if disconnecting the tube assembly (cable L3), before starting...
  • Page 412: Checking The Height Adjustment

    10 Maintenance Sirona Dental Systems GmbH 10.2 Checking the height adjustment Service Manual ORTHOPHOS SL 10.2 Checking the height adjustment Check the threaded rod and motor for ➢ Perform a visual inspection of height adjustment motor (A) and abrasion spindle (B) for abrasion.
  • Page 413: Checking The Forehead And Temple Supports

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.3 Checking the forehead and temple supports Check whether the height adjustment limit switches are functioning properly ➢ Manually press the actuators (C) of both limit switches (D) one after the other while the height adjustment motor is running. The motor must stop.
  • Page 414: Check Rotation Unit On The Sensor Unit For Ease Of Access

    10 Maintenance Sirona Dental Systems GmbH 10.4 Check rotation unit on the sensor unit for ease of access Service Manual ORTHOPHOS SL 10.4 Check rotation unit on the sensor unit for ease of access ✔ The unit must be switched on. 1. To switch from the PAN or CEPH program group to the 3D program group (or the other way around), touch the icons at the top edge of the touchscreen.
  • Page 415: Inspecting The Bite Block Holder

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.6 Inspecting the bite block holder 10.6 Inspecting the bite block holder * only for units with an occlusal bite block Check whether the bite block, contact ➢ Insert the bite block, contact segment and chin rest one after the...
  • Page 416: Checking The Ceph Sensor Holder

    10 Maintenance Sirona Dental Systems GmbH 10.7 Checking the ceph sensor holder Service Manual ORTHOPHOS SL 10.7 Checking the ceph sensor holder Check whether the sensor locks and unlocks easily and is firmly seated in the holder 1. Carefully insert the sensor upward into the holder: The sensor must snap in place audibly.
  • Page 417: Checking The Laser Light Localizers

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.8 Checking the laser light localizers 10.8 Checking the laser light localizers CAUTION Risk of injury to eyes. The unit contains lasers of Class 1. Keep a distance of at least 4" (10 cm) between eye and laser. Do not look into the laser beam.
  • Page 418 10 Maintenance Sirona Dental Systems GmbH 10.8 Checking the laser light localizers Service Manual ORTHOPHOS SL Checking the PAN laser beam 1. Affix a piece of white cardboard between the temple supports. 2. Rotate the mirror by pressing into the left depression (D) of the toolbar.
  • Page 419 Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.8 Checking the laser light localizers 1. Select the different volume programs one after the other with different collimations and measure the distance of the light beams. Minimum and maximum distances apply between the laser localizers.
  • Page 420: Checking The X-Ray Images

    10 Maintenance Sirona Dental Systems GmbH 10.9 Checking the X-ray images Service Manual ORTHOPHOS SL 10.9 Checking the X-ray images The check must be performed for pan, ceph and 3D X-ray images. For 2D X-ray images only: Check whether the X-ray images taken...
  • Page 421: Checking The Tube Data

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.10 Checking the tube data 10.10 Checking the tube data 10.10.1 Checking the tube voltage Preparing the measurement 1. Attach the Mult-O-Meter sensor in the middle of the sensor (on 2D sensor side).
  • Page 422: Checking The Tube Current

    10 Maintenance Sirona Dental Systems GmbH 10.10 Checking the tube data Service Manual ORTHOPHOS SL 10.10.2 Checking the tube current NOTICE Damage to the measuring unit The ring assembly and the tube assembly move during the measurement. Make sure that the measuring wires are sufficiently long to allow for the ring movement and that the measuring unit is in a secure position so that it will not fall down.
  • Page 423 Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.10 Checking the tube data 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 424 10 Maintenance Sirona Dental Systems GmbH 10.10 Checking the tube data Service Manual ORTHOPHOS SL 4. DANGER! After switching off the unit, wait at least 4 minutes (LED V500 on the DX6 must no longer be on) before removing jumper H from the DX6 board.
  • Page 425: Checking The Fan And Temperature Sensor In The Tube Assembly

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.10 Checking the tube data Concluding the measurement 1. Switch the unit on via switch (A) (see also Operating Instructions). 2. Loosen the screw (D) and remove the lid of the tube assembly cover (E).
  • Page 426: Checking The Radiation Time

    10 Maintenance Sirona Dental Systems GmbH 10.10 Checking the tube data Service Manual ORTHOPHOS SL 10.10.4 Checking the radiation time Preparing the measurement 1. Attach the Mult-O-Meter sensor in the middle of the sensor unit (DCS- side). 2. Set the main switch (A) to I (see also Operating Instructions).
  • Page 427: Test Exposures/Test Images

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images 10.11 Test exposures/Test images The 2D/3D test exposures and test images are used for fast checking of the unit adjustment and calibration. "Control exposures" menu The following test exposures are available: "Pan - diaphragm control exposure"...
  • Page 428: Calling "Test Exposures" Menu

    10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL 10.11.1 Calling "Test exposures" menu ✔ SIDEXIS 4 is installed. 1. Start SIDEXIS 4. "ORTHOPHOS SL" "Quality assurance" 2. In SIDEXIS 4, call the configuration menu. Configuration menu "ORTHOPHOS SL" /"Quality assurance"...
  • Page 429: Test Exposures

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images 10.11.2 2D test exposures Pan - diaphragm test exposure (2D) 10.11.2.1 "Pan - diaphragm control exposure" You can use the to check the adjustment of the Pan - diaphragm adjustment [ → 187].
  • Page 430: Pan - Symmetrical Test Exposure (2D)

    10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL Pan - Symmetrical test exposure (2D) 10.11.2.2 "Pan - symmetrie control exposure" You can use the to check the adjustment of the Pan - symmetry adjustment [ → 193].
  • Page 431: Pan - Quality Test Exposure (2D)

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images Pan - quality test exposure (2D) 10.11.2.3 "Pan - quality control exposure" You can use the to create a panoramic exposure with full rotation. This function enables you to simulate a quality test, for example, in a similar way to an acceptance/constancy test.
  • Page 432 10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL "Start acquisition" 6. Click on the button. Exposure readiness is established. A dialog window displays the status of readiness to exposure. The service routine used for the corresponding exposure is displayed on the control panel, along with the specific exposure parameters.
  • Page 433: Pan - Test Image (2D)

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images Pan - test image (2D) 10.11.2.4 "Pan - test image" You can use the menu to check the 2D data path of the unit and the operability of the DCS sensor.
  • Page 434 10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 435: Pan - Dark Current Exposure (2D)

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images Pan - dark current exposure (2D) 10.11.2.5 "Pan - dark current exposure" You can use the menu to check the DCS sensor and the data path. "Control exposures"...
  • Page 436 10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 437: Ceph - Secondary Diaphragm (2D)

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images Ceph - secondary diaphragm (2D) 10.11.2.6 "Ceph - secondary diaphragm" You can use the menu to check the adjustment of the Ceph secondary diaphragm. "Control exposures" 1. Call up the menu [ →...
  • Page 438: Ceph - Quality Test Exposure (2D)

    10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL Ceph - quality test exposure (2D) 10.11.2.7 "Ceph - quality control exposure" You can use the to create a Ceph exposure with full rotation. This function enables you to simulate a quality test, for example, in a similar way to an acceptance/constancy test.
  • Page 439 Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images 5. Secure the contrast element using a hygienic cap. 6. Attach the ceph test phantom to the clip provided for that purpose. 7. Push the ceph test phantom with the clip from below on to the secondary diaphragm of the ceph arm.
  • Page 440 10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL "Control exposures" 9. Call the menu [ → 428]. "2D control exposures" "Ceph 10. In the structure tree, under , select the - quality control exposure" element. "Ceph - quality In the action area, the menu appears to create the control exposure"...
  • Page 441: Ceph - Test Image (2D)

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images Ceph - test image (2D) 10.11.2.8 "Ceph - test image" You can use the menu to check the 2D data path of the unit and the operability of the Ceph sensor.
  • Page 442 10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 443: Test Exposures

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images 10.11.3 3D test exposures Diaphragm test exposure (open) (3D) 10.11.3.1 "Diaphragm control exposure (open)" You can use the to check the adjustment of the 3D diaphragm for overexposure when the diaphragm is open.
  • Page 444 10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 445: Diaphragm Test Exposure (Collimated) (3D)

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images Diaphragm test exposure (collimated) (3D) 10.11.3.2 "Diaphragm control exposure (collimated)" You can use the to adjust the Pan diaphragm with reduced diaphragm opening (to check dose waste on all sides).
  • Page 446 10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 447: Diaphragm Test Exposure (Standard) (3D)

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images Diaphragm test exposure (standard) (3D) 10.11.3.3 "Diaphragm control exposure (standard)" You can use the to check the adjustment of the original 3D diaphragm to check the circulating edge.
  • Page 448 10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL The diaphragm is correctly adjusted if a black surrounding edge can be detected. 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 449: Quality Test Exposure (3D)

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images Quality test exposure (3D) 10.11.3.4 "Quality control exposure Sirona" You can use the to create a volume exposure with a full rotation. This function enables you to simulate a quality test, for example, in a similar way to an acceptance/constancy test.
  • Page 450 10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL 5. Press the release button. Press and hold down the button until the exposure is complete, the preview image is displayed in the exposure window, and the acoustic signal that indicates the end of the exposure (double beep) sounds (if it has been configured).
  • Page 451: Quality Test Exposure Din 6868-161 (3D)

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images Quality test exposure DIN 6868-161 (3D) 10.11.3.5 "Quality control exposure DIN 6868-161" You can use the to create a volume exposure with a full rotation according to DIN 8686-161. This function enables you to simulate a quality test, for example, in a similar way to an acceptance/constancy test.
  • Page 452: Test Image (3D)

    10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL 7. Press the release button. Press and hold down the button until the exposure is complete, the preview image is displayed in the exposure window, and the acoustic signal that indicates the end of the exposure (double beep) sounds (if it has been configured).
  • Page 453 Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images 4. Press the release button. Press and hold down the button until the exposure is complete, the preview image is displayed in the exposure window, and the acoustic signal that indicates the end of the exposure (double beep) sounds (if configured).
  • Page 454: Dosimetry (3D)

    10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL Dosimetry (3D) 10.11.3.7 A dosimeter for pulsed radiation (e.g. Mult-O-Meter 512L) is required for dosimetry. "Dose measurement Sirona" You can use the menu to perform dosimetry independently of the unit adjustment and calibration.
  • Page 455: Dosimetry Din 6868-161 (3D)

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images Dosimetry DIN 6868-161 (3D) 10.11.3.8 A dosimeter for pulsed radiation (e.g. Mult-O-Meter 512L) is required for dosimetry. "Dose measurement DIN 6868-161" You can use the menu to perform dosimetry according to DIN 6868-161 independently of the unit adjustment or calibration.
  • Page 456: Dark Current Exposure (3D)

    10 Maintenance Sirona Dental Systems GmbH 10.11 Test exposures/Test images Service Manual ORTHOPHOS SL 6. Press the release button. Press and hold down the button until the exposure is complete, the preview image is displayed in the exposure window, and the acoustic signal that indicates the end of the exposure (double beep) sounds (if configured).
  • Page 457 Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.11 Test exposures/Test images 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 458: Checking The Cables For Damage

    10 Maintenance Sirona Dental Systems GmbH 10.12 Checking the cables for damage Service Manual ORTHOPHOS SL 10.12 Checking the cables for damage Check whether the cables feeding the ➢ Perform a visual inspection of the power cable, protective ground unit are OK wire, control cables and data cables.
  • Page 459: Checking The Idling Rollers

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.13 Checking the idling rollers 10.13 Checking the idling rollers Check whether the idling rollers (A) are ➢ Manually turn the ring (B) and check it for smooth and easy movement. If the ring does not move smoothly and easily: Remove the housing covers and check the idle rollers (A) for dirt and foreign particles.
  • Page 460: Checking The Grounding Straps

    10 Maintenance Sirona Dental Systems GmbH 10.14 Checking the grounding straps Service Manual ORTHOPHOS SL 10.14 Checking the grounding straps Grounding strap in the stand Check whether the grounding straps 1. Perform a visual and “hands-on” inspection of the grounding straps to...
  • Page 461: Checking The Cable Shields

    Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.15 Checking the cable shields 10.15 Checking the cable shields Shield on the tube assembly Shield on the housing Check whether the cable shielding is OK ➢ Perform a visual and “hands-on” inspection of the cable shields to ensure that they have complete and firm contact at the positions marked.
  • Page 462: Checking The Protective Ground Wires

    10 Maintenance Sirona Dental Systems GmbH 10.16 Checking the protective ground wires Service Manual ORTHOPHOS SL 10.16 Checking the protective ground wires DANGER General - Ground wire test Perilous shock hazard! It is essential to switch the unit off and to wait at least one more 1 minute before beginning the check! 1.
  • Page 463 Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.16 Checking the protective ground wires 1. Connect the power source between the measuring points specified in the table for at least 5 s. 2. Measure the voltage drop with the voltmeter, measure the current with the ammeter, and calculate the resistance using the formula R = U / I.
  • Page 464 10 Maintenance Sirona Dental Systems GmbH 10.16 Checking the protective ground wires Service Manual ORTHOPHOS SL Measuring point A: Central protective ground wire 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 465 Sirona Dental Systems GmbH 10 Maintenance Service Manual ORTHOPHOS SL 10.16 Checking the protective ground wires Measuring points B and C: GNYE power connection and 2nd ground Checking measuring points B and C - protective ground wire wire Power cable to the unit...
  • Page 466: Checking The Device Leakage Current

    DIN VDE 0100-710 applies in Germany. For measurements, Sirona recommends an automatic tester (example Automatic tester in accordance with IEC 62353 illustration) which complies with standard IEC 62353. If you do not use an automatic tester, please pay attention to the specifications in the standard IEC 62353.
  • Page 467: Dismantling And Disposal

    Sirona Dental Systems GmbH 11 Dismantling and disposal Service Manual ORTHOPHOS SL 11.1 Dismantling and reinstallation Dismantling and disposal 11.1 Dismantling and reinstallation When dismantling and reinstalling the system, proceed according to the installation instructions for new installation in order to guarantee its proper functioning and stability.
  • Page 468 11 Dismantling and disposal Sirona Dental Systems GmbH 11.2 Disposal Service Manual ORTHOPHOS SL Abroad: For country-specific information on disposal, contact your local dental dealers. The X-ray tube assembly for this product contains an X-ray tube with a X-ray tubes and counterweights potential implosion hazard, a small amount of beryllium, a lead lining and mineral oil.
  • Page 469: Service Manual History

    Sirona Dental Systems GmbH 12 Service Manual History Service Manual ORTHOPHOS SL Service Manual History Version 1: New creation Version 2: PC software and device firmware updated, chapter “Updating the unit firmware [ → 68]” added, chapter “Sensor unit [ → 378]” updated,...
  • Page 470 12 Service Manual History Sirona Dental Systems GmbH Service Manual ORTHOPHOS SL 64 95 258 D3632 D3632.076.01.03.02 09.2016...
  • Page 472 tÉ=êÉëÉêîÉ=íÜÉ=êáÖÜí=íç=ã~âÉ=~åó=~äíÉê~íáçåë=ïÜáÅÜ=ã~ó=ÄÉ=êÉèìáêÉÇ=ÇìÉ=íç=íÉÅÜåáÅ~ä=áãéêçîÉãÉåíëK «=páêçå~=aÉåí~ä=póëíÉãë=dãÄe=OMNS péê~ÅÜÉW ÉåÖäáëÅÜ mêáåíÉÇ=áå=dÉêã~åó aPSPOKMTSKMNKMPKMO MVKOMNS ûKJkêKW= NOO=UMT páêçå~=aÉåí~ä=póëíÉãë=dãÄe `çåí~Åí=áå=íÜÉ=rp^W páêçå~=aÉåí~äI=fåÅK SQ=VR=ORU=aPSPO c~Äêáâëíê~≈É=PN QUPR=páêçå~=aêáîÉ lêÇÉê=kç aJSQSOR=_ÉåëÜÉáã `Ü~êäçííÉI=k`=OUOTP dÉêã~åó ïïïKëáêçå~KÅçã...

This manual is also suitable for:

Orthophos sl 3dOrthophos sl 3d ceph

Table of Contents