9 Repair
9.12 Sensor unit
384
Connecting a sensor unit
9.12.2.2
1. Insert leads L83A and L83B from the rotation unit, as shown in the
illustration, into plugs X100, X700 and X800 on board DX83.
If a Flat Panel Detector is present:
2.
Insert lead L84 from the Flat Panel Detector, as shown in the
illustration, into plugs X601 and X602 on board DX83.
3. Lay the leads as shown in the illustration.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
A
D3632.076.01.03.02 09.2016
64 95 258 D3632