Location 23: Flat Panel Detector - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Error code
Description
E6 23 01
Error in initializing the digital flat
panel detector.
Error code
Description
E6 23 03
Malfunction in communication with
flat panel detector.
Error code
Description
E6 23 10
No image data obtained from flat
panel detector or interruption in data
flow.
Error code
Description
E6 23 15
Serial number of flat panel detector
could not be read out.
Error code
Description
E7 23 16
File operation malfunction in flat
panel detector mode.
64 95 258 D3632
D3632.076.01.03.02 09.2016
5.4.11

Location 23: Flat Panel Detector

Actions required
E6 23 01
● Check the connection of the flat panel
Actions required
E6 23 03
● Check wiring and plug connections from
● Replace sensor unit, if necessary.
Actions required
E6 23 10
● Check wiring and plug connections from
● Replace sensor unit, if necessary.
Actions required
E6 23 15
If the error does
errors ...
● Replace sensor unit.
Actions required
E6 23 16
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch on the unit.
If the error occurs repeatedly ...
● Replace sensor unit.
detector to board DX83.
flat panel detector (DX83 (X101) to board
DX1 (X500) and that the DCS sensor is
securely attached.
flat panel detector (DX83 (X101) to board
DX1 (X500) and that the DCS sensor is
securely attached.
not
occur with other causal
5 Messages
5.4 List of error messages
see
see
S. [ → 30]
S. [ → 378]
see
S. [ → 30]
S. [ → 378]
see
S. [ → 378]
see
S. [ → 378]
119

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 3dOrthophos sl 3d ceph

Table of Contents