8 Service routines
8.1 Perform service routines via the control panel
264

S005: Test step 7

8.1.6.5
Configuring continuous operation of the tube assembly fan
NOTICE
Contamination of the fan and the single tank housing
Because of the expected greater contamination of the fan and of the
single tank housing, the configuration of continuous operation is
documented irreversibly on the tube assembly in order to obtain
additional information in the case of returned goods.
IMPORTANT
The tube assembly fan can be configured for continuous operation at
the express wish of the customer This can be done by running this
service routine. The customer should be informed in advance about the
increased noise generation to be expected. A demonstration of the fan
noise can be made with service routine .
Selection field
1
* factory settings
1. Call service routine S005.7 [ → 249].
The operating mode of the fan is displayed in selection field 1.
2. To set the fan to continuous operation, use the arrow keys (A) to
select the code "00" (see table).
The Memory key (B) lights up.
3. Save the setting [ → 251].
4. Exit the service routine [ → 252].
Service Manual ORTHOPHOS SL
Code
Function
00
Fan switches off (controlled operation)*
01
Fan switches to continuous operation
Sirona Dental Systems GmbH
64 95 258 D3632
D3632.076.01.03.02 09.2016