10 Maintenance
10.11 Test exposures/Test images
454

Dosimetry (3D)

10.11.3.7
A dosimeter for pulsed radiation (e.g. Mult-O-Meter 512L) is required for
dosimetry.
You can use the
independently of the unit adjustment and calibration.
1. Attach the Mult-O-Meter sensor in the middle of the sensor unit.
2. Select a program from the 3D program group on the control panel.
The unit moves to the 3D exposure position.
3. Call up the
4. Select the
"3D control exposures"
under
In the action area, the menu appears to create the
measurement Sirona"
5. Click on the
Exposure readiness is established. A dialog window displays the
status of readiness to exposure.
The service routine used for the corresponding exposure is
displayed on the control panel, along with the specific exposure
parameters.
6. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal that indicates the end of the
exposure (double beep) sounds (if configured).
7. Read off the dose from the Mult-O-Meter. The dose value must be
between 1.2 and 2.3 mGy.
8. Remove the Mult-O-Meter from the sensor unit again.
"Dose measurement Sirona"
"Control exposures"
menu [ → 428].
"Dose measurement Sirona"
.
.
"Start acquisition"
button.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
menu to perform dosimetry
element in the structure tree
"Dose
64 95 258 D3632
D3632.076.01.03.02 09.2016