Adjusting The Laser Area Display (Vol) - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

9 Repair
9.7 Laser light localizers
A
364

Adjusting the laser area display (VOL)

9.7.4.2
1. Switch the unit on via the switch (A) (see also Operating Instructions).
B
C
2. Press the R key.
3. Touch the 3D symbol at the top of the touchscreen.
4. Press the R key.
5. Touch the light localizer key on the control panel.
The X-ray radiation indicator (B) lights up briefly.
After approx. 2 seconds, the green LED (C) in the upper part of
the control panel lights up. This LED remains lit as long as the
unit is on.
The start screen appears on the touchscreen of the Easypad and
the system's self-adjustment routine starts running (for approx. 1
minute). The diaphragm moves into position. Forehead and
temple supports on the panoramic unit are moved to the home
position.
On completion of the self-adjustment routine, the main menu
appears on the touchscreen. Help message H301 prompts you to
move the unit into the starting position.
The unit moves to its starting position.
The 3D program group is selected.
The diaphragm and the sensor move into the starting position for
3D exposures.
The laser line of the first laser module (count direction from
bottom) depicts the sensor unit on the cover (flat panel detector
side). A distance of 5 mm should be measured between the laser
line and joint of the sensor cover.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016