2. Use the arrow keys (A) to select the required kV/mA level and the
required radiation time (see table).
3. Initiate the radiation.
Radiation is released from any tube assembly position.
There is no rotational movement.
The primary diaphragm is opened fully.
The radiation time corresponds to the selected radiation time.
Important: If you let go of the release button before this radiation
time expires, radiation will be terminated prematurely. The actual
radiation time is not displayed.
When you release radiation during the cool-down interval, a
countdown of the remaining waiting time is displayed in the title
bar of the control panel (automatic exposure blocking).
4. Exit the service routine [ → 252].
8.1 Perform service routines via the control panel
8 Service routines
257