Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
SR*
S007
S007.1
S007.2
64 95 258 D3632
D3632.076.01.03.02 09.2016
8.1.7

S007: Fault buffer

SA**
Function
Fault buffer
No
View error memory
Yes
Clear error memory
* SR=service routine, ** SA=security access
8.1.7.1

S007: Test step 1

View error memory
Besides the service routine S007.1, you may also query [ → 59] the error
memory using the "Unit information".
Symbol on the control
panel
Patient symbol key 1 (B)
Patient symbol key 2 (C)
Patient symbol key 3 (D)
Selection field
1
2
3
* factory settings
1. Call service routine S007.1 [ → 249].
2. Use the patient symbol keys (B, C, D) to select the step width for
scrolling between the error events (see table).
3. Use the arrow keys (A) in selection field 1 to select the required error
event (in the example 1).
4. Exit the service routine [ → 252].
8.1 Perform service routines via the control panel
Status
is selected
is selected
is selected
Selection/display
Error event
Error code for the selected event
Date and time of the selected
error event
The selected patient symbol key lights up.
The corresponding error message is displayed [ → 84] in
selection field 2.
Selection field 3 displays the date and time of the error event.
8 Service routines
Function
Step width for scrolling
between error events = 1*
Step width for scrolling
between error events = 10
Step width for scrolling
between error events = 100
265