6 Troubleshooting
6.9 Error analysis of X-RAY control signal path
E3 13 40 occurs after the unit is switched
on:
Error code
Description
E3 13 40
Short circuit in signal path between
board DX11 and release button A2
during power-on.
E6 13 41
Release signal missing on board
DX11 at start of exposure.
E3 13 42
Short circuit in signal path between
board DX11 and release button A2
during operation of the unit.
158
without installed remote control
Error messages
with and without installed remote control
Error messages
Actions required
● Replace cable L117 or L108.
● Replace board DX1.
● Replace board DX11.
● Replace cable L117 or L108.
● Replace board DX1.
● Replace board DX11.
● Replace cable L117 or L108.
● Replace board DX1.
● Replace board DX11.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
see
S. [ → 404],
S. [ → 396]
S. [ → 404],
S. [ → 396]
S. [ → 404],
S. [ → 396]
64 95 258 D3632
D3632.076.01.03.02 09.2016