Service Manual ORTHOPHOS SL
Error code
Description
E6 06 05
Data transfer error or dialog error to
bootloader of module
Only occurs in connection with
software update.
Error code
Description
E6 06 06
Module failed in TTP (detected on
master side)
This error may also occur in
connection with other causal error
messages! Please also observe the
causal error message! It appears
only after you acknowledge the first
error message.
TTP = Time Trigger Protocol
Error code
Description
E6 06 07
TTP-Timeout error (detected on
slave side).
The module was temporarily not
addressed by the master:
● Undervoltage on the master
side
● Procedure error in the software
● Master (DX11) receives no
return commanding from the
module
This error may also occur in
connection with other causal error
messages! Please also observe the
causal error message! It appears
only after you acknowledge the first
error message.
TTP = Time Trigger Protocol
Error code
Description
E6 06 08
General fault detected locally on
module (slave side). CAN controller
being reinitialized.
64 95 258 D3632
D3632.076.01.03.02 09.2016
Actions required
E6 06 05
● Repeat the software update.
● Check the CAN bus.
If the error occurs repeatedly or the module
is no longer addressable ...
● Replace the tube assembly.
Actions required
E6 06 06
● Check the CAN bus.
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
Actions required
E6 06 07
● Check the CAN bus.
● Check power supply of board DX11;
measuring point 3.3 V on board DX1
(see wiring diagrams)
If 3.3 V is present ...
● Replace board DX11.
If 3.3 V is not present ...
● Replace board DX1.
● Check cable L6, replace if necessary. S. [ → 146]
● Check tube assembly (DX6), replace if
necessary.
Actions required
E6 06 08
● Check the CAN bus.
● Check software versions on the info
screen or by running service routine
S008.2, perform software update if
necessary.
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
● Replace the tube assembly.
5 Messages
5.4 List of error messages
see
S. [ → 68]
S. [ → 141]
S. [ → 372]
see
S. [ → 141]
S. [ → 68]
see
S. [ → 141]
S. [ → 395]
S. [ → 395]
S. [ → 400]
S. [ → 372]
see
S. [ → 141]
S. [ → 267],
S. [ → 68]
S. [ → 68]
S. [ → 372]
85