S012: Can Bus Service; S012: Test Step 1 - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

8 Service routines
8.1 Perform service routines via the control panel
SR*
S012
S012.1
S012.2
S012.3
S012.4
Checking the module
Clearing the counter for the module
272
8.1.11

S012: CAN bus service

SA**
Function
CAN bus service
No
Presence display of modules
No
Inquiry of the CAN status register of the modules
No
Resetting the CAN status register of the modules
No
Display of CAN bus cycle on the LEDs of the modules
* SR=service routine, ** SA=security access
IMPORTANT
The CAN bus service is not yet implemented for the module DX11!
8.1.11.1

S012: Test step 1

Presence display of modules
Selection field
1
2
➢ Call service routine S012.1 [ → 249].
➢ Use the arrow keys (A) in selection field 1 to select the required
module.
1. To delete the counter, press the T key (B).
2. Exit the service routine [ → 252].
Parameter/Display
Subunit
● Counter value of
CAN bus events
● Presence code behind
the counter value:
P = module present
L = module lost
The counter value of the CAN bus events (change from "P" to "L"
since the last switch-on of the unit) of the selected module
processed so far is displayed in selection field 2 with the
presence code of the module ("L" or "P") (see table).
Once the module has been selected, the T key (B) lights up.
The counter is then reset to "0".
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Range of
values
DX6 - DX91
64 95 258 D3632
D3632.076.01.03.02 09.2016