Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
"Service"
In the main menu
64 95 258 D3632
D3632.076.01.03.02 09.2016
8.2

Running service routines via the web service

NOTE: The web service is only available in English
NOTE: The following service routines can be run using the Sirona Web
service:
● S008.2: Overview of module software versions
● S008.3: Confirming the unit serial number
● S017.2: Configuring the hardware version
● S017.3: Enter the country group code
● S017.4: Select a language
● S017.6: Activating / deactivating the remote control
● S017.8: Selecting the kV/mA level series
● S017.13/S017.14: Enable/disable the welcome screen or the lines in
the welcome screen
● S017.15: Activate/deactivate
the acoustic signal for end of exposure
● S017.18: Activation/deactivation of occlusal bite block function
● S037: Manual input of static network settings (IP address, default
gateway address, and subnet mask)
The unit and computer are turned on.
The unit is logged into the network as an X-ray component.
1. Start an Internet browser such as Internet Explorer or Firefox.
2. In the address line, enter "https://" and the IP address of your unit.
Example: https://192.168.15.240
NOTICE! The IP address of your unit can be found on the Info screen.
3. Confirm your input with the "Enter" key.
A security prompt appears.
8.2 Running service routines via the web service
8 Service routines
327