Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
A
B
4. Loosen the 4 screws (B) and unhinge the DCS sensor from the lugs
(A) of the rotation unit.
5. Install the new sensors in the reverse order of removal. For the
instructions on connecting the sensor unit, see chapter "Connecting
a sensor unit [ → 384]".
6. Reattach the covers of the sensor unit (D and E) with the 2 screws (F)
in reverse order of removal.
Only if there is a flat panel detector:
7.
Restore the previously installed sensor data using the menu
/ Restore sensor data"
the unit (refer to chapter "Restoring sensor data [ → 74]").
8. Perform a complete unit adjustment or calibration.
"Restore sensor data"
⇒
9 Repair
9.12 Sensor unit
"Backup
on board DX83 of
383