S014: Rotation motor service

SA**
Function
Rotation motor service
No
Free travel of rotational drive
No
Display of light barrier signals of rotational drive
* SR=service routine, ** SA=security access
8.1.12.1

S014: Test step 3

Free travel of rotational drive
This service routine implements the free travel of the rotational drive in
both directions.
1. Call service routine S014.3 [ → 249].
2. Touch the arrow keys (A) to move the ring to the right (+) or to the left
(–). The ring keeps moving as long as you touch the key or until it
reaches the respective end position.
3. Exit the service routine [ → 252].

S014: Test step 4

8.1.12.2
Display of light barrier signals of rotational drive
This service routine implements display of the switching state of rotation
motor light barrier V1_3.
1. Call service routine S014.4 [ → 249].
2. Turn the ring manually. Patient symbol key 1 (B) lights up to indicate
that the light barrier has been released.
3. Exit the service routine [ → 252].
8.1 Perform service routines via the control panel
Once the service routine has been selected, the arrow keys (A)
are displayed in selection field 1.
Once the service routine has been selected, both the Service key
and patient symbol key 1 (B) are displayed on the control panel.
8 Service routines
275