9 Checking the system adjustment

9.1 Test exposures

9.1 Test exposures
1.
9.1.1 Test exposure menu
The panoramic X-ray unit is already completely adjusted
on delivery. Check the adjustment using the TEST
EXPOSURE menu to make sure that the unit was not inad -
vertently deadjusted during transport.
The TEST EXPOSURE menu is started from SIDEXIS XG:
UTILITIES
CONSTANCY TEST
-
X
RAY DEVICE
SERVICE EXPOSURE
-
X
RAY COMPONENT
DIAPHRAGM TEST EXPOSURE
IMPORTANT
-
The SELECT X
RAY DEVICE and SELECT X
NENT dialogs appear only if several x-ray devices or x-ray
components are available for selection.
IMPORTANT
When you open the TEST EXPOSURE menu, the unit
switches from the user mode to the PC service mode
logged by the PC. The service mode is indicated by the
SERVICE display on the Multipad (2.).
In the PC service mode, the control options
that are available on the Multipad are determined by SI -
DEXIS and the currently selected service routine. General
control of the unit by means of the Multipad (as in the user
mode) is not possible in this mode.
The test exposures are used to check the adjustment. They
can be started without a password.
84
XCXP
SELECT
SELECT
-
-
RAY COMPO
Installation Instructions ORTHOPHOS XG 3
2.
S E R V I C E
You can switch between the PAN DIAPHRAGM and PAN
SYMMETRY menus by clicking the corresponding tab with
the mouse. To quit the TEST EXPOSURE menu, click
CANCEL .
Displays on the Multipad
During the adjustment procedure, different service routines
are started from SIDEXIS; they are displayed one after the
other on the Multipad.
If there are any error messages displayed during adjust -
ment, please follow the instructions provided in the Service
Manual.