1. Select the different volume programs one after the other with different
collimations and measure the distance of the light beams.
Minimum and maximum distances apply between the laser
localizers.
Replace any light localizer that does not respond as desired.
2. Press on the light localizer key again.
The light localizer is switched back off again.
Checking the horizontal laser beam ceph (FH)
Tip: When checking or adjusting the light localizer, you may use a PA or
AP position on the ear plugs to assess the light beam.
1. Touch the light localizer key (B) on the touchscreen.
The light localizers are switched on.
2. Check the position of the horizontal laser beam at the ceph.
The laser beam must run horizontally at the level of the ear plug
position between the template supports. If this is not the case,
adjust the laser localizer.
3. Press on the light localizer key again.
The light localizer is switched back off again.
10 Maintenance
10.8 Checking the laser light localizers