7 Unit adjustment and calibration
7.1 General information on unit adjustment and calibration
176
7.1.6
Test phantom for adjustment and calibration
Needle phantom for panoramic adjustment
7.1.6.1
NOTICE
XGPF-36318; Use needle phantom correctly
Risk of damage to unit
Make absolutely sure that the needle phantom is inserted into the bite
block holder of the unit in such a way that the front needle (B) of the
phantom points
figure). Otherwise, this can result in a collision between the sensor and
the needle phantom and thereby damage the unit.
For
"Sensor adjustment"
adjustments, the needle phantom (A) must be inserted into the pan bite
block holder of the unit. For
adjustment, the needle phantom must be taken out of the bite block
holder.
away from the unit in the direction of the patient
and
"Symmetrie adjustment"
"Diaphragm adjustment"
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
(see
B
A
panorama
panorama
64 95 258 D3632
D3632.076.01.03.02 09.2016