8 Service routines
8.1 Perform service routines via the control panel
294
2. Use the arrow keys (H) in selection field 1 to select the required
program group.
3. Use the arrow keys (I) to select the required nominal angle (possible
range: -17° to +22°).
Tip: You can also set the required angle manually on the occlusal bite
block and transfer this angle to selection field 2 by pressing the T key
(J).
By pressing patient symbol key 1 (G) you can activate and deactivate
the laser light localizer. You can use the laser light as a reference for
the angle adjustment. When the laser light localizer is activated, the
patient symbol key 1 is lit.
The Memory key (B) lights up.
4. To save the nominal value, touch the Memory key (B) (R key (C)
lights up) followed by the R key (C).
5. Exit the service routine.