Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
SR*
S011
S011.14
S011.24
64 95 258 D3632
D3632.076.01.03.02 09.2016
8.1.10
S011: Dosimetry (without ring movement)
Unit is radiating X-rays.
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
SA**
Function
Dosimetry (without ring movement)
yes
Dosimetry with continuous radiation
yes
Dosimetry with pulsed radiation
* SR=service routine, ** SA=security access
8.1.10.1
S011: Test step 14
Dosimetry with continuous radiation
A standard dosimeter is required for dosimetry.
Selection field
1
2
1. Attach the Mult-O-Meter sensor in the middle of the sensor unit (flat
panel detector side).
2. Call service routine S011.14 [ → 249].
8.1 Perform service routines via the control panel
WARNING
Parameters
kV/mA levels
Radiation time
(not selectable)
8 Service routines
Range of values
85kV/5mA (3.2s)
85kV/7mA (5s)
85kV/10mA (5s)
85kV/13mA (5s)
3.2 s
5 s
269