Service Manual ORTHOPHOS SL
Error code
Description
E6 06 56
Error during auto-compensation.
Error code
Description
E6 06 60
TDI signal from board DX11 to
board DX6 is disturbed.
TDI = Signal to start synchronized
readout sequence and to prepare
the next exposure
Error code
Description
E6 06 65
Tube current or tube voltage is too
high in standby mode.
Error code
Description
E6 06 66
Impermissible tube type.
Error code
Description
E6 06 67
Light guide input TDI is active during
switch-on.
TDI = Signal to start synchronized
readout sequence and to prepare
the next exposure
64 95 258 D3632
D3632.076.01.03.02 09.2016
Actions required
E6 06 56
● Please contact the Sirona Customer
Service Center (CSC) to find out
whether a bug fix by means of a
software update is possible and
perform such an update if necessary.
● Let the tube assembly cool down for
approx. 30 mins and repeat this
procedure.
If the error occurs repeatedly ...
● Replace the tube assembly.
Actions required
E6 06 60
● Replace cable L15.
● Replace board DX1.
● Replace the tube assembly.
Actions required
E6 06 65
● Replace the tube assembly.
Actions required
E6 06 66
● Check tube type of tube assembly
using extended detail query, replace
tube assembly if necessary.
Actions required
E6 06 67
● Check TDI signal:
1. Switch off the unit.
2. Disconnect cable L15 at board DX11.
3. Switch unit on.
4. Perform visual check at socket J5:
● If light is visible: Replace board DX11. S. [ → 395]
● If no light is visible: Replace the tube
assembly.
5 Messages
5.4 List of error messages
see
S. [ → 68]
S. [ → 372]
see
S. [ → 406],
S. [ → 395],
S. [ → 372]
see
S. [ → 372]
see
S. [ → 372]
see
S. [ → 372]
89