7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
Horizontal correction
Vertical correction
224
1. Loosen the screws (A) slightly (approx. 2 - 3 turns).
Tip: Do not unscrew them fully!
2. Adjust the ear plugs in the horizontal direction by turning the screw
(B) counterclockwise or clockwise. The direction of rotation for left-
arm (C) and right-arm (D) versions is shown in the drawing below.
3. Tighten the screw (A).
1. Turn the rotary table counterclockwise approx. 100 degrees until you
can see the screw (E) through the opening in the cover plate.
2. Loosen the screws (E) slightly (approx. 2 - 3 turns).
Tip: Do not unscrew them fully!
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016