Cephalometer; Replacing The Ceph Sensor Holder; What Has To Be Done After Replacing The Ceph Sensor - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
C
B
A
E
64 95 258 D3632
D3632.076.01.03.02 09.2016
9.13

Cephalometer

9.13.1

Replacing the ceph sensor holder

1. Remove the sensor (A).
2. NOTICE! Do not tear off cable L35.
Loosen the 3 screws (B) and remove the connection socket (C).
3. Detach the cable shield from terminal (D) and the connector screw
connection (loosen the two screws (E) from the rear) and thread the
connector to the rear through the connection socket.
4. NOTICE! Do not forget to reattach the cable shield.
Install the new sensor holder in reverse order of the disassembly.
D
What has to be done after replacing the ceph sensor holder?
9.13.1.1
➢ Perform the ceph unit adjustment (see section ("Unit adjustment and
calibration [ → 160]").
9 Repair
9.13 Cephalometer
391