Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
Pan symmetry adjustment
7.2.1.3
7.2.1.3.1
Automatic adjustment: Pan - Symmetry
✔
The needle phantom is in the bite block holder on the unit [ → 176].
1. In the structure tree, under
adjustment"
The
2. Establish [ → 175] receptivity.
3. Create an exposure (60 kV / 3 mA; 14.1 s) [ → 175].
The adjustment values S1, S2 and S3 are automatically
determined from the exposure and entered in the text boxes of
"Symmetrie adjustment"
the
The exposure is displayed in the exposure window.
7.2 Adjustment and calibration via the calibration menu
"Pan"
, click on the
element (S010.2).
"Pan - Symmetrie"
menu is displayed in the action area.
menu.
7 Unit adjustment and calibration
"Symmetrie
193