S015: Actuator Service; S015: Test Step 5 - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

8 Service routines
8.1 Perform service routines via the control panel
SR*
S015
S015.5
276
8.1.13

S015: Actuator service

SA**
Function
Actuator service
No
Functional test of actuators 1 + 2
* SR=service routine, ** SA=security access
8.1.13.1

S015: Test step 5

Testing of actuators 1 and 2
The use of this service routine is suitable for troubleshooting motor and
cable defects. Furthermore, the possibility of performing isolated travel
with an individual drive makes it possible to release the clamping of the
actuator/rotation system and check the stiffness of the individual
adjusting axes.
Symbol on the control
panel
Patient symbol key 1 (C)
Patient symbol key 2 (D)
Patient symbol key 1+2
(C+D)
1. Call service routine S015.5 [ → 249].
2. Touch the arrow keys (A) in selection field 1 to move actuator 1 to the
right (+) or to the left (–). The actuator movement continues as long
as the key is pressed.
The R key (E) can be used to move the system (actuator 1 and 2)
back to the pan home position. Patient symbol keys 1 and 2 light up
in this position.
3. Touch the arrow keys (B) in selection field 2 to move actuator 2 to the
right (+) or to the left (–). The actuator movement continues as long
as the key is pressed.
The R key (E) can be used to move the system (actuator 1 and 2)
back to the pan home position. Patient symbol keys 1 and 2 light up
in this position.
4. Exit the service routine [ → 252].
Status
lit
not lit
lit
not lit
lit
Once the service routine has been selected, the arrow keys
(A+B) are displayed in selection field 1 and 2.
The patient symbol key 1 (C) shows the switching state of the
light barrier of actuator 1, the patient symbol key 2 (D) shows the
switching state of the light barrier of actuator 2 (see table).
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Meaning
Light barrier of actuator 1 is
located in the switching flag
Light barrier of actuator 1 is not
located in the switching flag
Light barrier of actuator 2 is
located in the switching flag
Light barrier of actuator 2 is not
located in the switching flag
Unit is in the pan home position
64 95 258 D3632
D3632.076.01.03.02 09.2016