Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Error code
Description
E5 81 77
Ceph: Voltage error detected on
sensor. (CCD voltages on board
DX81C).
If this error occurs in connection with
other causal error messages (e.g.
E6 61 60, E6 91 50), take
appropriate action.
Error code
Description
E5 81 78
Ceph: Invalid data at start of
exposure.
64 95 258 D3632
D3632.076.01.03.02 09.2016
Actions required
E5 81 77
● Check cable L35/L38, replace if
necessary.
● Check connector X503 on board DX1,
replace board DX1 if necessary.
● Run software update.
● Check ceph sensor holder, replace if
necessary.
● Check ceph sensor (replacement sensor),
replace if necessary.
Actions required
E5 81 78
● Acknowledge error and repeat procedure.
If the error occurs repeatedly ...
● Run software update.
● Check the version compatibility of the unit
software and SIDEXIS 4, perform software
update if necessary.
● Seek advice from the Sirona Customer
Service Center (CSC) with regard to the
settings/sensor information in the Xab.ini
file.
5 Messages
5.4 List of error messages
see
S. [ → 153],
S. [ → 404],
S. [ → 396],
S. [ → 68],
S. [ → 391]
see
S. [ → 68]
133