Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
IMPORTANT
When installing a new DX11 which already has the same software
status as the overall system, you must nevertheless perform a software
update to this status, so that an administrative entry can be written to the
memory of the DX11.
7. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
Error message E61505 (undefined unit serial number) is
displayed.
8. Call the service menu [ → 248].
9. Call up service routine S008.3 [ → 249].
10. Check the serial number
IMPORTANT: The unit serial number is located on the nameplate of
the unit.
NOTICE! If the serial number is incorrect, cancel the update and
contact the Sirona Customer Service Center!
11. Confirm the serial number using service routine S008.3 [ → 267].
12. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
13. Call the service menu [ → 248].
14. Call service routine S017 [ → 249] and perform the unit configuration
using service routines S017.2 - S017.18 [ → 277].
15. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
16. If the travel height of the unit has to be limited:
Set the travel height with service routine S018.2 [ → 289].
17. Perform a software update to the current software version.
All modules are updated in accordance with the configuration.
The process is completed.