Pan sensor calibration (DCS)

7.2.1.4
NOTE: For sensor calibration, 7 exposures are taken (for the different kV/
mA combinations). The exposure is only displayed in the exposure
window at the end of the calibration procedure.
1. Open the calibration menu [ → 164].
2. In the structure tree under
(DCS)"
element (S010.10).
IMPORTANT
"Sensor calibration (DCS)"
For the
(for the applicable kV/mA combinations). After taking the first exposure,
receptivity is automatically established again by the unit. This is
repeated until the last exposure has been taken.
3. Establish [ → 175] receptivity.
4. Take the exposure for the first kV/mA combination (60 kV / 5 mA ;
15.1s) [ → 175].
After taking the exposure, the unit automatically re-establishes
receptivity.
5. Take the exposure for the second Kv/mA combination [ → 175].
After taking the exposure, the unit automatically re-establishes
receptivity.
7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
"Pan"
"Sensor calibration
, click on the
, a total of 7 exposures must be taken