7 Unit adjustment and calibration
7.1 General information on unit adjustment and calibration
Configuration menu "ORTHOPHOS SL" /"Device administration"
164
7.1.2

Calibration menu

The menu guides you through the procedure to adjust and calibrate the
unit.
7.1.2.1

Opening the calibration menu

IT package is installed.
is installed.
1. Start SIDEXIS 4.
2. In SIDEXIS 4, call the
configuration menu.
3. Select the desired X-ray component from the unit list.
4. Click on the
A dialog box to enter the service password opens.
5. Enter the service password (see section "Password protection") and
confirm by pressing
The calibration menu opens.
"ORTHOPHOS SL"
"Device calibration"
button.
"OK"
.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
"Device administration"
/
64 95 258 D3632
D3632.076.01.03.02 09.2016