7 Unit adjustment and calibration
7.2 Adjustment and calibration via the calibration menu
182
7.2.1.1.1
Automatic adjustment: Panoramic sensor (standard)
The needle phantom is in the bite block holder on the unit [ → 176].
1. Open the calibration menu [ → 164].
2. At the bottom of the structure tree, click on the
element (S010.1) under
The
The menu supports precision adjustment and coarse adjustment
(precision adjustment is pre-selected). Perform a precision
adjustment first. In most cases, previous coarse adjustment is not
necessary. [ → 170]
3. Establish receptivity [ → 175].
4. Create an exposure (60 kV / 3 mA ; 0.7 s). [ → 175]
The adjustment values S1, S2 and S3 are automatically
determined from the exposure and entered in the text boxes of
"Sensor adjustment"
the
The exposure is displayed in the exposure window.
"Pan"
.
"Sensor adjustment"
menu is displayed in the action area.
menu.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
"Sensor adjustment"
64 95 258 D3632
D3632.076.01.03.02 09.2016