Operation; Setting The Acquisition System To 3D Camera; General - Sirona Cerec 3 Operating Instructions Manual

Acquisition unit
Hide thumbs Also See for Cerec 3:
Table of Contents

Advertisement

7 Operation

Setting the acquisition system to 3D camera

28
7
Operation
7.1
Setting the acquisition system to 3D
camera
In order to use the 3D camera, the acquisition must be set to "3D
camera".
Select the command "Settings" / "Configuration" /
1.
"Acquisition system" in the menu line.
The "Configuration" dialog box appears.
Select "3D camera" and confirm with "OK".
2.
The 3D camera will remain selected until you select "Scanner" or
"inEos".
7.2

General

Aligning the 3D camera
The direction of acquisition must
coincide with the insertion axis of
the preparation prepared by the
dentist.
right
Depth of focus and focusing
The telecentric optics, which cause objects to be displayed with a constant
size regardless of how far away from the prism they are, have a depth of focus
which is sufficient to capture deep preparations.
The image definition is determined by the distance between the 3D camera
and the preparation.
Sirona Dental Systems GmbH
Operating Instructions for the Acquisition Unit CEREC 3
If the camera is held at an oblique
angle to the prepared insertion axis,
the wall closer to the lens will be
registered with an undercut; the wall
further away from the lens will be
fully displayed, thus causing the
occlusal margin angle to be
presented unfavorably there and
obstructing the automatic margin
detection.
wrong
D 3344.201.01.25.23 01.2011
63 22 700 D 3344