9.6

Control panel

9.6.1

Replacing the user interface

1. Press into slit (A) of the housing cover with a screwdriver (do not pry!)
and remove the defective user interface from the control panel.
2. Pull cables L9 (gray) and L10 (green) off connectors X102 (L9) and
X103 (L10) on board DX7 of the defective user interface.
3. Plug the cables into connectors X102 (L9) and X103 (L10) of board
DX7 on the new user interface.
4. CAUTION! Ensure that the flat-ribbon cable (B) is not crushed,
Clip the new user interface onto the control panel.
A
9 Repair
9.6 Control panel
349