Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
10.11.3
3D test exposures
Diaphragm test exposure (open) (3D)
10.11.3.1
You can use the
adjustment of the 3D diaphragm for overexposure when the diaphragm is
open.
1. Call up the menu
2. Select the
structure tree under
In the action area, the menu appears to create the
control exposure (open)"
3. Click on the
Exposure readiness is established. A dialog window displays the
status of readiness to exposure.
The service routine used for the corresponding exposure is
displayed on the control panel, along with the specific exposure
parameters.
4. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal that indicates the end of the
exposure (double beep) sounds (if configured).
The exposure is displayed in the exposure window.
See below for reference image.
No foreign bodies may appear on the image.
"Diaphragm control exposure (open)"
"Control exposures"
"Diaphragm control exposure (open)"
"3D control exposures"
.
"Start acquisition"
button.
10 Maintenance
10.11 Test exposures/Test images
to check the
[ → 428].
element in the
.
"Diaphragm
443