Adjusting the earplug alignment
7.2.1.9
7.2.1.9.1
Checking the ear plug alignment
✔
The ceph sensor is plugged into the ceph slot on the unit.
1. In the structure tree, under
earplug alignment"
The
action area.
The unit is moved into the ceph exposure position.
2. Move the ear plug holders completely apart and swing them into the
beam direction.
7.2 Adjustment and calibration via the calibration menu
"Ceph"
element (S010.17).
"Adjustment of earplug alignment"
7 Unit adjustment and calibration
"Adjustment of
, click on the
menu is displayed in the
221