Sirona ORTHOPHOS SL 2D Service Manual page 123

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
Error code
Description
E3 42 30
R key pressed during power-on.
Error code
Description
E3 42 31
Release button pressed during
power-on.
The hardware signal for radiation
release is applied on board DX42
when the unit is switched on.
64 95 258 D3632
D3632.076.01.03.02 09.2016
Actions required
E3 42 30
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch on the unit, making sure that the
remote control is not pressed during
booting.
4. Repeat procedure and check function.
If the error occurs repeatedly ...
● Replace remote control, see Installation
Instructions.
Actions required
E3 42 31
● See section entitled Error analysis of X-
RAY control signal path [ → 156].
5 Messages
5.4 List of error messages
see
see
S. [ → 156]
123