Advertisement

4 General operating procedures
4.1 Switching the unit on
A
48
IMPORTANT
After switch-on, the unit requires approx. 1 min. of warm-up time.
At the same time, the self-adjustment routine for the mechanics and the
electronics of the unit is performed. If a key is pressed during the self-
adjustment routine, an error message will display on the control panel.
1. Turn the main switch (A) to position I.
2. Wait for approx. 1 minute.
B
C
3. Check whether the patient symbols on the touchscreen can be
selected in exactly the right position.
If problems occur during selection, adjust the touchscreen [ → 236].
4.1.1

Factory setting after switch-on

The unit has the following factory configuration on delivery:
● The acoustic signal for end of exposure is activated.
● English is preconfigured as the unit language.
● The preview image is switched on.
● The welcome screen is switched on.
● The first name, last name and date of birth lines are displayed on the
welcome screen.
● Ambient light (background lighting) is set to "white".
If the customer requires a different configuration, this can be implemented
via service routine S017 or the Web service (ambient light).
The radiation indicator X-ray (C) lights up for approx. one second
as a functional check.
For a number of seconds the height adjustment keys light up
blue, then quickly change to green.
After approx. 2 seconds, the green LED B in the upper part of the
control panel lights up. This LED remains lit as long as the unit is
on.
The start screen is displayed on the touchscreen for several
seconds.
The program selection is then displayed on the touchscreen.
The forehead support and temple supports are completely open.

Sirona Dental Systems GmbH

Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016