9 Repair
9.6 Control panel
352
What has to be done after replacing the control panel?
9.6.2.1
IMPORTANT
Use the most current software version available on the
SL Firmware DVD"
1. Switch the unit on.
2. Perform a software update to the current unit software version.
3. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
4. Check the control panel and the display elements for correct
functioning: When the unit is switched on, all of the display elements
must light up briefly.
9.6.3
Laying cables for control panel replacement
A
Strain relief
Service Manual ORTHOPHOS SL
.
Sirona Dental Systems GmbH
"ORTHOPHOS
64 95 258 D3632
D3632.076.01.03.02 09.2016