Important Information For Repacking And Transport - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
4.8.3

Important information for repacking and transport

IMPORTANT
If a used carton on which one of the shockwatch or tiltwatch indicators
has already been tripped is used to package the unit, please make an
entry to that effect on the delivery note.
Packing the cephalometer
1. Call service routine S034.6 [ → 316].
After calling the service routine, an inactive progress indicator is
displayed in selection field 1. Selection field 2 shows the
characters "FFFF".
2. Touch the T key (A).
The cephalometer moves to the packing position. The procedure
is visualized by an active progress indicator in selection field 1.
When the cephalometer has reached its packing position, "0000"
is displayed in selection field 2.
3. Exit the service routine. [ → 252]
Packing the cephalometer
IMPORTANT
There are markings on the column of the unit. The bottom edge of the
slide cover must be at the same height as the markings in the column.
DANGER
Shock hazard!
Be sure to switch off the line power supply before connecting the line
voltage!
1. Switch the unit on and move it to its packing height by actuating the
up/down keys on the control panel.
Bite block height = 965 mm (displayed as height on the control
panel)
Bottom edge of slide cover = 702 mm
2. Pack the panoramic X-ray unit
(For packing condition, refer to the "Delivery" section of the
installation manual).
Attaching the transport safety device
➢ Install the transport safety device by following the same procedure as
Dismantling in reverse order.
4 General operating procedures
4.8 Using demo mode – operation without radiation release
79