4.6.2

Backing up sensor data

You can use the
Backing up sensor data1
(factory calibration of the sensor) from the DX83 board and save it as a
Zip file.
NOTE: As the sensor data is stored on the DX83 board, this is necessary,
for example, before changing the DX83 board. The stored sensor data
must then be restored on the new DX83 board after replacing the board.
This can be done [ → 74] via the menu
"Backup / Restore sensor data"
The
Save sensor data2
1. In the structure tree, click on the
The
"Backup sensor data"
2. Click on the
"Download Backup"
The
A selection dialog box opens in your browser.
3. In the selection dialog box, click the
select the central image data directory
(...\PDATA\sensordata) as the storage location.
4. Confirm the selection by clicking the
The sensor data is downloaded from the DX83 board and saved
as a Zip file in the central image data directory
IMPORTANT
Sensor data must be stored in the central image data directory
If no selection dialog box appears in your browser, copy the Zip file
manually from the download directory of your browser to the central
image data directory
4.6 Backing up / restoring / installing sensor data
"Backup sensor data"
menu to load the sensor data
"Restore sensor data"
menu is opened.
"Backup sensor data"
menu is displayed.
"Backup sensor data"
button.
menu opens.
"Save"
"OK"
"PDATA"
.
4 General operating procedures
.
element.
"Save as"
/
button and
"PDATA"
button.
"PDATA"
.
"PDATA"
73