S033: Test Of Ceph Image Path (For Ceph Units Only); S033: Test Step 10 - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
SR*
S033
S033.10
64 95 258 D3632
D3632.076.01.03.02 09.2016
8.1.19

S033: Test of ceph image path (for ceph units only)

SA**
Function
Test of the ceph image path
No
Test of the ceph image path
* SR=service routine, ** SA=security access
8.1.19.1

S033: Test step 10

Signal tests for sensor in ceph slot
Selection field
1
1. Call service routine S033.10 [ → 249].
2. Use the arrow keys (A) in selection field 1 to select the mode of the
required synchronized readout sequence (see table).
IMPORTANT
If parameter "HSIC4x4M" is selected:
If the R key should light up first, the sensor holder must first be moved
to the corresponding position. To do so, press the R key.
3. Touch the T key (B).
8.1 Perform service routines via the control panel
Parameters
HSIC4x4
HSIC4x4M
After test step 10 is selected, selection field 1 displays the current
selection for synchronized readout sequence.
Once the mode for the synchronized readout sequence has been
selected, the T key (B) lights up.
The synchronized readout sequence is activated.
8 Service routines
Function
Sensor in ceph mode with
synchronized readout sequence 4x4
clocked out without ceph scan
Sensor in ceph mode with
synchronized readout sequence 4x4
clocked out with ceph scan
307