4 General operating procedures
4.6 Backing up / restoring / installing sensor data
74
4.6.3
Restoring sensor data
Previously saved sensor data (factory calibration of the sensor) can be
restored on the DX83 board using the
IMPORTANT
After the sensor data has been restored, the unit adjustment and
calibration must be performed [ → 160] in full.
If no complete unit adjustment and calibration has been performed, no
patient images are possible.
NOTE: As the sensor data is stored on the DX83 board, this is necessary,
for example, after changing the DX83 board. To that end, the sensor data
must have been backed up [ → 73] before replacing the DX83 board using
"Backup sensor data"
the
✔
"Backup / Restore sensor data"
The
1. In the structure tree, click on the
"Restore sensor data"
The
2. Click on the
directory (refer to chapter "Backing up sensor data [ → 73]").
3. Click on the
The sensor data stored in the Zip file is restored on the DX83
board.
4. Perform a complete unit adjustment and calibration [ → 160].
"Restore sensor data"
menu.
menu is opened.
"Restore sensor data"
menu is displayed.
"Select File"
button and select the desired Zip file in the
"Restore"
button.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
menu.
element.
64 95 258 D3632
D3632.076.01.03.02 09.2016