Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016

S018: Test step 4

8.1.15.3
Check the height adjustment sensor system
This service routine is used to move the unit up or down as far as the limit
switches using the Up/Down keys on the control panel. The "soft limit
positions" set by the software are ignored in this case.
Display on the control
panel
Patient symbol key 1
Patient symbol key 2
Patient symbol key 3
1. Call service routine S018.4.
Once the service routine has been selected, the current height
position is displayed in selection field 1.
Patient symbol keys 1 to 3 (F) show the switching state of the limit
switches (see table)
If the RESPECTIVE patient symbol key is lit, the corresponding
switch is activated.
2. Use the UP/DOWN keys on the control panel to move the unit up and
down and use the patient symbol keys (F) to check the switching
states.
3. Exit the service routine.
8.1 Perform service routines via the control panel
Status
Meaning
lit
Correction switch activated
not lit
Correction switch not
activated
lit
Lower limit switch activated
not lit
Lower limit switch not
activated
lit
Upper limit switch activated
not lit
Upper limit switch not
activated
8 Service routines